タイトル | Imaging of Am-241 in the lung by using an imaging plate | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_c94f | |||||
資源タイプ | conference object | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Hirota, Masahiro
× Hirota, Masahiro× et.al× 廣田 昌大 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | A new Am-241 lung monitoring system without shielding was devised by using an imaging plate system. The Lawrence Livermore National Laboratory's realistic torso phantom containing an Am-241 lung was covered by imaging plates sealed in lightproof bags. The imaging plate system displayed Am-241 lung images characteristic of the lung shape of the torso phantom. The imaging plate system's lower detection limits of 14 Bq for 60 min exposure and 6 Bq for 300 min were the same levels as those of the phoswich detectors and the germanium detectors placed in shielded rooms. The imaging plate system for 60 min exposure detected about 2% of the annual limit of 740 Bq for Am-241 inhalation. A lung monitoring system using imaging plates is applicable for Am-241 lung monitoring. | |||||
会議概要(会議名, 開催地, 会期, 主催者等) | ||||||
内容記述タイプ | Other | |||||
内容記述 | The Second Asian and Oceanic Congress for Radiation Protection(第2回アジア・オセアニア放射線防護会議) | |||||
発表年月日 | ||||||
日付 | 2006-10-13 | |||||
日付タイプ | Issued |