Measurements of Dose Distributions in Lateral Directions in Water Using CR-39 for Therapeutic Carbon-Ion Beams

https://repo.qst.go.jp/records/61056
https://repo.qst.go.jp/records/61056
a26b8ab9-f67c-451c-9949-7d051200f67c
アイテムタイプ 会議発表用資料 / Presentation(1)
公開日 2005-08-02
タイトル
タイトル Measurements of Dose Distributions in Lateral Directions in Water Using CR-39 for Therapeutic Carbon-Ion Beams
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference output
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Kohno, Ryosuke

× Kohno, Ryosuke

WEKO 605039

Kohno, Ryosuke

Search repository
河野 良介

× 河野 良介

WEKO 605040

en 河野 良介

Search repository
抄録
内容記述タイプ Abstract
内容記述 Purpose: In order to develop more precise dose averaged LET calculation algorithm on the basis of the pencil beam algorithm (PBA) for carbon ion radiotherapy, we have to calculate depth and lateral dose averaged LET distribution for the pencil beam. Particularly, since the lateral dose averaged LET distribution is formed by scattered particles and fragment particles, we expected the distribution is somewhat complex. In this study, we focused on the lateral dose averaged LET distribution. Lateral dose averaged LET distributions were measured using CR-39 LET detector and analyzed.
Method and Materials: Carbon ion irradiation was carried out using the horizontal beam line at Heavy Ion Medical Accelerator in Chiba. A therapeutic carbon beam with energies of 290 MeV per nucleon (MeV/n) was used. Using a CR-39 solid state track detector, lateral dose averaged LET distributions in water were measured at various depths for collimated mono-energetic beam.
Results: We analyzed etch pits on the CR-39, and lateral dose averaged LET distributions in the region of penumbra were obtained. As a result, their LETs in even penumbra region were about constant at each depth.
Conclusion: Lateral dose averaged LET distributions in water were measured at various depths for collimated mono-energetic beam using a CR-39 solid state track detector. It was obvious that their LETs in even penumbra region about constant at each depth. We found that lateral dose averaged LET can be calculated roughly using depth dose averaged LET for broad beam.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述タイプ Other
内容記述 AAPM 47th Annual Meeting
発表年月日
日付 2004-07-28
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 21:46:44.015348
Show All versions

Share

Share
tweet

Cite as

Other

print

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX
  • ZIP

コミュニティ

確認

確認

確認


Powered by WEKO3


Powered by WEKO3