Development Of Radial Echo Planar Imaging with Central High-Resolution Area ( REPI with CHRA )

https://repo.qst.go.jp/records/59248
https://repo.qst.go.jp/records/59248
ecd0515c-7d41-4864-bbf8-ad1a3fe97fcf
Item type 会議発表用資料 / Presentation(1)
公開日 2002-06-13
タイトル
タイトル Development Of Radial Echo Planar Imaging with Central High-Resolution Area ( REPI with CHRA )
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference object
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Yoneyama, Ryuichi

× Yoneyama, Ryuichi

WEKO 589720

Yoneyama, Ryuichi

Search repository
Kandatsu, Susumu

× Kandatsu, Susumu

WEKO 589721

Kandatsu, Susumu

Search repository
米山 隆一

× 米山 隆一

WEKO 589722

en 米山 隆一

Search repository
神立 進

× 神立 進

WEKO 589723

en 神立 進

Search repository
抄録
内容記述タイプ Abstract
内容記述 Purpose
The most challenging aspect of imaging dynamically changing objects by MRI is achieving the desired temporal resolution without sacrificing the required spacial resolution. The present authors have developed a new MRI sequence termed Radial Echo Planer Imaging with a Central High-Resolution Area (REPI with CHRA), in which high temporal resolution is balanced with high spatial resolution in the center of the field of view (FOV).
\nMethod
Using a 1.5T Sigma Horizon LX MRI system (GE Medical Systems, Milwaukee, WI, U.S.A.) and EPIC (Environment for Programming in C), we developed the REPI with CHRA technique, in which radii of a sector area of k-space are scanned in one excitation, with the first radius being twice the width of the subsequent radii. The entire k-space is scanned in multiple excitations. After all radii other than the first in a sector area are zero-filled, the image is reconstructed by back projection. A MRI fluoroscopy of the heart and a contrast enhanced MR angiography (CEMRA) of the head were taken by applying REPI with CHRA continuously.
\nResults
Using 16-shot REPI with CHRA, with a FOV of 24 cm, we achieved a TR of 320 ms for the acquisition of a 128 ´ 128 image with a resolution of 1 mm in the center of the FOV and a resolution of 2 mm in the peripheral area. As REPI with CHRA is basically a radial sampling method, the image suffered less from the phase error artifacts characteristic of multi-shot EPI. Using 16-shot REPI with CHRA, we took images of the heart and CEMRA of the head both with a FOV of 24 cm, for 32 s. Using the view-sharing technique, we obtained images with a temporal resolution of 20 ms and a spatial resolution of 1 mm in the center area. The images exhibited fewer density discontinuities than by ordinary EPI. The motion artifact manifested as a star-like shadow, which offers greater clarity than the blurring characteristic of ordinary EPI.
\nConclusion
We proved that REPI with CHRA can be used successfully to balance high temporal resolution with high spatial resolution, which is necessary for imaging dynamically changing objects. The images produced using the technique exhibit minimal density discontinuity, and less obtrusive motion artifacts. These characteristics render this technique suitable for imaging moving objects or for CEMRA.
発表年月日
日付 2001-11-30
日付タイプ Issued
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 22:07:03.586508
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3