Development of the X’tal Cube PET Detector With Segments of (0.77 mm)3
https://repo.qst.go.jp/records/49525
https://repo.qst.go.jp/records/49525fb482be2-a59a-43ec-92dc-eebc92be2211
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2019-01-31 | |||||
タイトル | ||||||
タイトル | Development of the X’tal Cube PET Detector With Segments of (0.77 mm)3 | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Nitta, Munetaka
× Nitta, Munetaka× Inadama, Naoko× Nishikido, Fumihiko× Yoshida, Eiji× Tashima, Hideaki× Kawai, Hideyuki× Yamaya, Taiga× Nitta, Munetaka× Inadama, Naoko× Nishikido, Fumihiko× Yoshida, Eiji× Tashima, Hideaki× Kawai, Hideyuki× Yamaya, Taiga |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | The depth-of-interaction detector which can identify gamma-ray detection position along the depth direction is essential for a PET scanner to reduce parallax error. The X’tal cube we have developed is composed of a scintillation crystal block segmented into cubes. Six arrays of multipixel photon counters (MPPCs) cover six surfaces of the crystal block. We developed the new X’tal cube consisting of 0.77 mm cubic segments. A (13.1 mm)3 LYSO crystal block had 3-D segmentation in a 17 × 17 × 17 array fabricated by the laser engraving technique. 4 × 4 MPPC arrays with a size of (13 mm)2 were used. Active area of each MPPC was (3.0 mm)2. To evaluate segment identification performance, we irradiated with gamma-rays from a 22Na source and obtained a 3-D position histogram made by the 3-D Anger-type calculation with all MPPC signals. The histogram showed 17 segments in a row can be identified clearly. Further, to measure its intrinsic spatial resolution with incident angles of 0◦ and 45◦, gamma-rays in a fan-beam were irradiated and scanned onto the X’tal cube. Spatial resolutions of 0.74 ± 0.10 mm and 0.89 ± 0.10 mm were obtained by both scan experiments, respectively. |
|||||
書誌情報 |
IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES 巻 2, 号 6, p. 564-573, 発行日 2018-11 |
|||||