ログイン
言語:

WEKO3

  • トップ
  • ランキング
To
lat lon distance
To

Field does not validate



インデックスリンク

インデックスツリー

メールアドレスを入力してください。

WEKO

One fine body…

WEKO

One fine body…

アイテム

  1. 原著論文

Evaluation of patient positional reproducibility on the treatment couch and its impact on dose distribution using rotating gantry system in scanned carbon-ion beam therapy

https://repo.qst.go.jp/records/49465
https://repo.qst.go.jp/records/49465
24423e88-c84b-4839-b5e0-3f2030e325f3
Item type 学術雑誌論文 / Journal Article(1)
公開日 2019-01-16
タイトル
タイトル Evaluation of patient positional reproducibility on the treatment couch and its impact on dose distribution using rotating gantry system in scanned carbon-ion beam therapy
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 金井, 貴幸

× 金井, 貴幸

WEKO 790667

金井, 貴幸

Search repository
古市, 渉

× 古市, 渉

WEKO 790668

古市, 渉

Search repository
森, 慎一郎

× 森, 慎一郎

WEKO 790669

森, 慎一郎

Search repository
Kanai, Takayuki

× Kanai, Takayuki

WEKO 790670

en Kanai, Takayuki

Search repository
Furuichi, Wataru

× Furuichi, Wataru

WEKO 790671

en Furuichi, Wataru

Search repository
Mori, Shinichiro

× Mori, Shinichiro

WEKO 790672

en Mori, Shinichiro

Search repository
抄録
内容記述タイプ Abstract
内容記述 Purpose: The daily variations in patient setup may cause beam range uncertainties. We evaluated the re- producibility of relative position between the patient and the treatment couch throughout the treatment course and assessed its effects on dose distributions when a beam passes through treatment couch using rotating gantry system.
Methods: We enrolled 1023 patients (=13072 fractions) treated by carbon-ion pencil beam scanning therapy. Seven treatment sites including prostate, head and neck, bone and soft tissue, rectum, liver, lung, and pancreas were investigated. Inter-fractional changes in couch position relative to the patient were defined as translational errors. Changes in couch rotation were defined as rotational errors. Treatment planning was performed for 4 patients in each of the treatment sites. Dose distributions were then re-calculated after the couch was shifted according to average, 95th percentile, and maximum values of translational error.
Results: Large positional errors (> 1.5 cm) were observed in 5% of treatment fractions. Positional errors were largest in prostate and pancreas patients, while smallest in head and neck and lung patients. There were no or only small changes in PTV-D95 and CTV-D95 values for almost all treatment sites. Clinically significant changes were observed in the duodenum (difference in D2cc values ranged from −55% to 28% with maximum couch shift) in pancreas treatment.
Conclusions: Although underdosage to the PTV or CTV was limited, significant overdoses to organs at risk were found. The improvement of immobilization technique and appropriate selection of gantry angles could reduce the uncertainties due to changes in patient position.
書誌情報 Physica medica

巻 57, p. 160-168, 発行日 2019-01
出版者
出版者 Elsevier
ISSN
収録物識別子タイプ ISSN
収録物識別子 1120-1797
DOI
識別子タイプ DOI
関連識別子 10.1016/j.ejmp.2018.12.013
関連サイト
識別子タイプ URI
関連識別子 https://www.physicamedica.com/article/S1120-1797(18)31349-8/pdf
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 18:42:06.006446
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3