Nakada, Yoshihiro
× Nakada, Yoshihiro× Okuda, Yasuo× Tsuge, Tatsuya× Suzuki, Jyunichi× Sakamoto, Hiroshi× Yamamoto, Tsuyoshi× Konishi, Yasuhiko× Tsujimoto, Takeshi× Nishiki, Shigeo× Satoh, Toshimitsu× Aoyama, Nobukazu× Morimoto, Kyohei× Aita, Masamichi× Yamashita, Yusuke× Yoshitake, Takayasu× Mukai, Masami× Yokooka, Yuki× Yokohama, Noriya× Akahane, Keiichi× Nakada, Yoshihiro× Okuda, Yasuo× Mukai, Masami× Yokooka, Yuki× Yokohama, Noriya× Akahane, Keiichi |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | The present work describes that we try to construct a system that collects dose information that performed CT examination from multiple facilities and unified management. The results of analysis are compared with other National diagnostic reference level (DRL), and the results are fed back to each facility and the cause of the abnormal value is investigated for dose optimization. Medical information collected 139 144 tests from 33 CT devices in 13 facilities. Although the DRL of this study is lower than that of Japan DRL, it was higher than the DRL of each country. When collecting all the examination, it is thought that the variation of the dose due to the error other than the intended imaging site is large. In future, we should continue to collect information in order to DRL renewal and we also think that it is desirable to collect information on physique and detailed scan region as well. | |||||
書誌情報 |
Radiation Protection Dosimetry 巻 181, 号 2, p. 156-167, 発行日 2018-02 |
|||||
出版者 | ||||||
出版者 | Oxford University Press | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0144-8420 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1093/rpd/ncy003 | |||||
関連サイト | ||||||
識別子タイプ | DOI | |||||
関連識別子 | https://doi.org/10.1093/rpd/ncy003 |