{"_buckets": {"deposit": "5fa3f2ff-1110-4b5e-88b3-8bf6dd37694b"}, "_deposit": {"created_by": 1, "id": "48110", "owners": [1], "pid": {"revision_id": 0, "type": "depid", "value": "48110"}, "status": "published"}, "_oai": {"id": "oai:repo.qst.go.jp:00048110", "sets": ["1"]}, "author_link": ["483102", "483100", "483105", "483103", "483098", "483104", "483101", "483099"], "item_8_biblio_info_7": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "2016-05", "bibliographicIssueDateType": "Issued"}, "bibliographicIssueNumber": "12", "bibliographicPageEnd": "2033", "bibliographicPageStart": "2028", "bibliographicVolumeNumber": "53", "bibliographic_titles": [{"bibliographic_title": "Journal of Nuclear Science and Technology"}]}]}, "item_8_description_5": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "A high temperature ratio (HTR) method has been proposed to correct the linear energy transfer (LET) dependence of thermoluminescent (TL) efficiency. However, reports in terms of application of the HTR method are concerned only with thermoluminescence detectors (TLDs) that offer point dose measurement. A slab-type TLD based on TL phosphor Li3B7O12:Cu has been developed and that has the potential to be a 3D dose verification tool. To realize the use of the slab-type TLD for heavy charged particle beam, the HTR method has been considered. To improve the reproducibility of HTR, the slow heating rate method is introduced in this report. The relation between TL-efficiency, HTR and LET for Li3B7O12:Cu was manifested and the TL-efficiency as a function of HTR was derived in an attempt to measure the absorbed dose without LET information. The feasibility of the HTR method in therapeutic carbon beams was evaluated by comparing the dose estimated by Li3B7O12:Cu and by an ionization chamber. The accuracy of dose estimation in carbon beams was improved by using the HTR method, but there is room for further improvement. The use of Li3B7O12:Cu in heavy charged particle beams can be materialized with further improvement of HTR sensitivity.", "subitem_description_type": "Abstract"}]}, "item_8_publisher_8": {"attribute_name": "出版者", "attribute_value_mlt": [{"subitem_publisher": "Taylor \u0026 Francis"}]}, "item_8_relation_14": {"attribute_name": "DOI", "attribute_value_mlt": [{"subitem_relation_type_id": {"subitem_relation_type_id_text": "10.1080/00223131.2016.1181010", "subitem_relation_type_select": "DOI"}}]}, "item_8_relation_17": {"attribute_name": "関連サイト", "attribute_value_mlt": [{"subitem_relation_name": [{"subitem_relation_name_text": "http://www.tandfonline.com/doi/full/10.1080/00223131.2016.1181010#.Vz5YdeT9jSc"}], "subitem_relation_type_id": {"subitem_relation_type_id_text": "http://www.tandfonline.com/doi/full/10.1080/00223131.2016.1181010#.Vz5YdeT9jSc", "subitem_relation_type_select": "URI"}}]}, "item_8_source_id_9": {"attribute_name": "ISSN", "attribute_value_mlt": [{"subitem_source_identifier": "0022-3131", "subitem_source_identifier_type": "ISSN"}]}, "item_access_right": {"attribute_name": "アクセス権", "attribute_value_mlt": [{"subitem_access_right": "metadata only access", "subitem_access_right_uri": "http://purl.org/coar/access_right/c_14cb"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "Chang, Weishan"}], "nameIdentifiers": [{"nameIdentifier": "483098", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "古場, 裕介"}], "nameIdentifiers": [{"nameIdentifier": "483099", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "福田, 茂一"}], "nameIdentifiers": [{"nameIdentifier": "483100", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Wakabayashi, Genichiro"}], "nameIdentifiers": [{"nameIdentifier": "483101", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Saitoh, Hidetoshi"}], "nameIdentifiers": [{"nameIdentifier": "483102", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Shinsho, Kiyomitsu"}], "nameIdentifiers": [{"nameIdentifier": "483103", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "古場 裕介", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "483104", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "福田 茂一", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "483105", "nameIdentifierScheme": "WEKO"}]}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "eng"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "journal article", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "Absorbed dose estimation using LET dependence in glow curve of thermoluminescent phosphor Li3B7O12:Cu in therapeutic carbon beams", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "Absorbed dose estimation using LET dependence in glow curve of thermoluminescent phosphor Li3B7O12:Cu in therapeutic carbon beams"}]}, "item_type_id": "8", "owner": "1", "path": ["1"], "permalink_uri": "https://repo.qst.go.jp/records/48110", "pubdate": {"attribute_name": "公開日", "attribute_value": "2017-07-20"}, "publish_date": "2017-07-20", "publish_status": "0", "recid": "48110", "relation": {}, "relation_version_is_last": true, "title": ["Absorbed dose estimation using LET dependence in glow curve of thermoluminescent phosphor Li3B7O12:Cu in therapeutic carbon beams"], "weko_shared_id": -1}
Absorbed dose estimation using LET dependence in glow curve of thermoluminescent phosphor Li3B7O12:Cu in therapeutic carbon beams
https://repo.qst.go.jp/records/48110
https://repo.qst.go.jp/records/48110e0b47ae0-f914-416e-9a5c-043c7c846e38
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2017-07-20 | |||||
タイトル | ||||||
タイトル | Absorbed dose estimation using LET dependence in glow curve of thermoluminescent phosphor Li3B7O12:Cu in therapeutic carbon beams | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Chang, Weishan
× Chang, Weishan× 古場, 裕介× 福田, 茂一× Wakabayashi, Genichiro× Saitoh, Hidetoshi× Shinsho, Kiyomitsu× 古場 裕介× 福田 茂一 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | A high temperature ratio (HTR) method has been proposed to correct the linear energy transfer (LET) dependence of thermoluminescent (TL) efficiency. However, reports in terms of application of the HTR method are concerned only with thermoluminescence detectors (TLDs) that offer point dose measurement. A slab-type TLD based on TL phosphor Li3B7O12:Cu has been developed and that has the potential to be a 3D dose verification tool. To realize the use of the slab-type TLD for heavy charged particle beam, the HTR method has been considered. To improve the reproducibility of HTR, the slow heating rate method is introduced in this report. The relation between TL-efficiency, HTR and LET for Li3B7O12:Cu was manifested and the TL-efficiency as a function of HTR was derived in an attempt to measure the absorbed dose without LET information. The feasibility of the HTR method in therapeutic carbon beams was evaluated by comparing the dose estimated by Li3B7O12:Cu and by an ionization chamber. The accuracy of dose estimation in carbon beams was improved by using the HTR method, but there is room for further improvement. The use of Li3B7O12:Cu in heavy charged particle beams can be materialized with further improvement of HTR sensitivity. | |||||
書誌情報 |
Journal of Nuclear Science and Technology 巻 53, 号 12, p. 2028-2033, 発行日 2016-05 |
|||||
出版者 | ||||||
出版者 | Taylor & Francis | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0022-3131 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1080/00223131.2016.1181010 | |||||
関連サイト | ||||||
識別子タイプ | URI | |||||
関連識別子 | http://www.tandfonline.com/doi/full/10.1080/00223131.2016.1181010#.Vz5YdeT9jSc | |||||