Absorbed dose estimation using LET dependence in glow curve of thermoluminescent phosphor Li3B7O12:Cu in therapeutic carbon beams

https://repo.qst.go.jp/records/48110
https://repo.qst.go.jp/records/48110
e0b47ae0-f914-416e-9a5c-043c7c846e38
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-07-20
タイトル
タイトル Absorbed dose estimation using LET dependence in glow curve of thermoluminescent phosphor Li3B7O12:Cu in therapeutic carbon beams
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Chang, Weishan

× Chang, Weishan

WEKO 483098

Chang, Weishan

Search repository
古場, 裕介

× 古場, 裕介

WEKO 483099

古場, 裕介

Search repository
福田, 茂一

× 福田, 茂一

WEKO 483100

福田, 茂一

Search repository
Wakabayashi, Genichiro

× Wakabayashi, Genichiro

WEKO 483101

Wakabayashi, Genichiro

Search repository
Saitoh, Hidetoshi

× Saitoh, Hidetoshi

WEKO 483102

Saitoh, Hidetoshi

Search repository
Shinsho, Kiyomitsu

× Shinsho, Kiyomitsu

WEKO 483103

Shinsho, Kiyomitsu

Search repository
古場 裕介

× 古場 裕介

WEKO 483104

en 古場 裕介

Search repository
福田 茂一

× 福田 茂一

WEKO 483105

en 福田 茂一

Search repository
抄録
内容記述タイプ Abstract
内容記述 A high temperature ratio (HTR) method has been proposed to correct the linear energy transfer (LET) dependence of thermoluminescent (TL) efficiency. However, reports in terms of application of the HTR method are concerned only with thermoluminescence detectors (TLDs) that offer point dose measurement. A slab-type TLD based on TL phosphor Li3B7O12:Cu has been developed and that has the potential to be a 3D dose verification tool. To realize the use of the slab-type TLD for heavy charged particle beam, the HTR method has been considered. To improve the reproducibility of HTR, the slow heating rate method is introduced in this report. The relation between TL-efficiency, HTR and LET for Li3B7O12:Cu was manifested and the TL-efficiency as a function of HTR was derived in an attempt to measure the absorbed dose without LET information. The feasibility of the HTR method in therapeutic carbon beams was evaluated by comparing the dose estimated by Li3B7O12:Cu and by an ionization chamber. The accuracy of dose estimation in carbon beams was improved by using the HTR method, but there is room for further improvement. The use of Li3B7O12:Cu in heavy charged particle beams can be materialized with further improvement of HTR sensitivity.
書誌情報 Journal of Nuclear Science and Technology

巻 53, 号 12, p. 2028-2033, 発行日 2016-05
出版者
出版者 Taylor & Francis
ISSN
収録物識別子タイプ ISSN
収録物識別子 0022-3131
DOI
識別子タイプ DOI
関連識別子 10.1080/00223131.2016.1181010
関連サイト
識別子タイプ URI
関連識別子 http://www.tandfonline.com/doi/full/10.1080/00223131.2016.1181010#.Vz5YdeT9jSc
関連名称 http://www.tandfonline.com/doi/full/10.1080/00223131.2016.1181010#.Vz5YdeT9jSc
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 23:31:51.403716
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3