Rapid phase-correlated rescanning irradiation improves treatment time in carbon-ion scanning beam treatment under irregular breathing

https://repo.qst.go.jp/records/48089
https://repo.qst.go.jp/records/48089
0023d731-b3ae-4f27-8484-41352e68f8f1
Item type 学術雑誌論文 / Journal Article(1)
公開日 2017-07-13
タイトル
タイトル Rapid phase-correlated rescanning irradiation improves treatment time in carbon-ion scanning beam treatment under irregular breathing
言語
言語 jpn
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 森, 慎一郎

× 森, 慎一郎

WEKO 482892

森, 慎一郎

Search repository
古川, 卓司

× 古川, 卓司

WEKO 482893

古川, 卓司

Search repository
森 慎一郎

× 森 慎一郎

WEKO 482894

en 森 慎一郎

Search repository
古川 卓司

× 古川 卓司

WEKO 482895

en 古川 卓司

Search repository
抄録
内容記述タイプ Abstract
内容記述 To shorten treatment time in pencil beam scanning irradiation, we developed rapid phase-controlled rescanning (rPCR), which irradiates two or more isoenergy layers in a single gating window. Here, we evaluated carbon-ion beam dose distribution with rapid and conventional PCR (cPCR).
4 dimensional computed tomography (4DCT) imaging was performed on 12 subjects with lung or liver tumors. To compensate for intrafractional range variation, the eld-speci c target volume (FTV) was calculated using 4DCT within the gating window (T20–T80). We applied an amplitude-based gating strategy, in which the beam is on when the tumor is within the gating window de ned by treatment planning. Dose distributions were calculated for layered phase-controlled rescanning under an irregular respiratory pattern, although a single 4DCT data set was used. The number of rescannings was eight times. The prescribed doses were 48 Gy(RBE)/1 fr (where RBE is relative biological effectiveness) delivered via four beam ports to the FTV for the lung cases and 45 Gy(RBE)/2 fr delivered via two beam ports to the FTV for the liver cases.
In the liver cases, the accumulated dose distributions showed an increased magnitude of hot/cold spots with rPCR compared with cPCR. The results of the dose assessment metrics for the cPCR and rPCR were very similar. The D95, Dmax, and Dmin values (cPCR/rPCR) averaged over all the patients were 96.3 ± 0.9%/96.0 ± 1.2%, 107.3 ± 3.6%/107.1 ± 2.9%, and 88.8 ± 3.2%/88.1 ± 3.1%, respectively. The treatment times in cPCR and rPCR were 110.7 s and 53.5 s, respectively.
rPCR preserved dose conformation under irregular respiratory motion and reduced the total treatment time compared with cPCR.
書誌情報 Physics in medicine and biology

巻 61, 号 10, p. 3857-3866, 発行日 2016-04
出版者
出版者 IOPscience
ISSN
収録物識別子タイプ ISSN
収録物識別子 0031-9155
DOI
識別子タイプ DOI
関連識別子 10.1088/0031-9155/61/10/3857
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 23:32:06.705581
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3