Range verification system using edge detection method for a scintillator and a CCD camera system
https://repo.qst.go.jp/records/47599
https://repo.qst.go.jp/records/47599b5ab1613-f349-43bc-86a8-79b7a86b7ec1
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2017-01-04 | |||||
タイトル | ||||||
タイトル | Range verification system using edge detection method for a scintillator and a CCD camera system | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Saotome, Naoya
× Saotome, Naoya× Furukawa, Takuji× Hara, Yousuke× Mizushima, Kota× Tansho, Ryohei× Saraya, Yuichi× Shirai, Toshiyuki× Noda, Koji× 早乙女 直也× 古川 卓司× 原 洋介× 水島 康太× 丹正 亮平× 皿谷 有一× 白井 敏之× 野田 耕司 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Purpose: Three-dimensional irradiation with a scanned carbon-ion beam has been performed from 2011 at our facility. We have been developed the rotating-gantry equipped with the scanning irradiation system. The number of combinations of beam properties to measure for the commissioning is more than 7200, i.e. 201 energy steps, 3 intensities, and 12 gantry angles. To compress the commissioning time, quick and simple range verification system is required. In this work, we develop quick range verification system using scintillator and CCD (charge-coupled device) camera and estimate the accuracy of the range verification. Methods: A cylindrical plastic scintillator block and a CCD camera were installed on the black box. The optical spatial-resolution of the system is 0.2 mm/pixel. The camera control system was connected and communicates with the measurement system that is part of the scanning system. The range was determined by image processing. Reference range for each energy beam was determined by a difference of Gaussian (DOG) method and the 80 percent of distal dose of the depth dose distribution that were measured by a large parallel-plate ionization chamber. We compared a threshold method and a DOG method. Results: We found that the edge detection method (i.e., the DOG method) is best for the range detection. The accuracy of range detection using this system is within 0.2 mm, and the reproducibility of same energy measurement is within 0.1 mm without setup error. Conclusions: The results of this study demonstrate that our range check system is capable of quick and easy range verification with sufficient accuracy. |
|||||
書誌情報 |
Medical Physics 巻 43, 号 4, p. 1754-1759, 発行日 2016-04 |
|||||
出版者 | ||||||
出版者 | American Association of Physicists in Medicine | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0094-2405 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1118/1.4943955 |