A proposal for PET/MRI attenuation correction with mu-values measured using a fixed-position radiation source and MRI segmentation
https://repo.qst.go.jp/records/47324
https://repo.qst.go.jp/records/473247f85218b-1e10-4bdb-b001-296d641fd802
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2015-11-13 | |||||
タイトル | ||||||
タイトル | A proposal for PET/MRI attenuation correction with mu-values measured using a fixed-position radiation source and MRI segmentation | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Kawaguchi, Hiroshi
× Kawaguchi, Hiroshi× Hirano, Yoshiyuki× Yoshida, Eiji× Kershaw, Jeffrey× Shiraishi, Takahiro× Suga, Mikio× Ikoma, Yoko× Obata, Takayuki× Ito, Hiroshi× Yamaya, Taiga× 川口 拓之× 平野 祥之× 吉田 英治× Kershaw Jeffrey× 白石 貴博× 菅 幹生× 生駒 洋子× 小畠 隆行× 伊藤 浩× 山谷 泰賀 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Several MRI-based attenuation correction methods have been reported for PET/MRI; these methods are expected to make efficient use of high-quality anatomical MRIs and reduce the radiation dose for PET/MRI scanning. The accuracy of the attenuation map (mu-map) from an MRI depends on the accuracy of tissue segmentation and the attenuation coefficients to be assigned (mu-values). In this study, we proposed an MRI-based mu-value estimation method with a non-rotational radiation source to construct a suitable mu-map for PET/MRI. The proposed method uses an accurately segmented tissue map, the partial path length of each tissue, and detected intensities of attenuated radiation from a fixed-position (rather than a rotating) radiation source to obtain the mu-map. We estimated the partial path length from a virtual blank scan of fixed-point radiation with the same scanner geometry using the known tissue map from MRI. The mu-values of every tissue were estimated by inverting a linear relationship involving the partial path lengths and measured radioactivity intensity. Validation of the proposed method was performed by calculating a fixed- point data set based upon real a real transmission scan. The root-mean-square error between the mu-values derived from a conventional transmission scan and those obtained with our proposed method were 2.4+/1.4%, 17.4+/-9.1% and 6.6+/-4.3% for brain, bone and soft tissue other than brain, respectively. Although the error estimates for bone and soft tissue are not insignificant, the method we propose is able to estimate the brain mu-value accurately and it is this factor that most strongly affects the quantitative value of PET images because of the large volumetric ratio of the brain. |
|||||
書誌情報 |
Nuclear Instruments & Methods in Physics Research Section A 巻 734, 号 Part B, p. 156-161, 発行日 2013-09 |
|||||