Detector to detector corrections: a comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams

https://repo.qst.go.jp/records/47165
https://repo.qst.go.jp/records/47165
7d0ffb1b-44d3-4a34-b85f-89269fae70bb
Item type 学術雑誌論文 / Journal Article(1)
公開日 2015-07-22
タイトル
タイトル Detector to detector corrections: a comprehensive experimental study of detector specific correction factors for beam output measurements for small radiotherapy beams
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Azangwe, G.

× Azangwe, G.

WEKO 471299

Azangwe, G.

Search repository
水野, 秀之

× 水野, 秀之

WEKO 471300

水野, 秀之

Search repository
福村, 明史

× 福村, 明史

WEKO 471301

福村, 明史

Search repository
矢島, 佳央理

× 矢島, 佳央理

WEKO 471302

矢島, 佳央理

Search repository
al., et

× al., et

WEKO 471303

al., et

Search repository
水野 秀之

× 水野 秀之

WEKO 471304

en 水野 秀之

Search repository
福村 明史

× 福村 明史

WEKO 471305

en 福村 明史

Search repository
矢島 佳央理

× 矢島 佳央理

WEKO 471306

en 矢島 佳央理

Search repository
抄録
内容記述タイプ Abstract
内容記述 Purpose: The aim of the present study is to provide a comprehensive set of detector specific correction factors for beam output measurements for small beams, for a wide range of real time and passive detectors. The detector specific correction factors determined in this study may be potentially useful as a reference data set for small beam dosimetry measurements.
Methods: Dose response of passive and real time detectors was investigated for small field sizes shaped with a micromultileaf collimator ranging from 0.6 × 0.6 cm2 to 4.2 × 4.2 cm2 and the measurements were extended to larger fields of up to 10 × 10 cm2. Measurements were performed at 5 cm depth, in a 6 MV photon beam. Detectors used included alanine, thermoluminescent dosimeters (TLDs), stereotactic diode, electron diode, photon diode, radiophotoluminescent dosimeters (RPLDs), radioluminescence detector based on carbon-doped aluminium oxide (Al2O3:C), organic plastic scintillators, diamond detectors, liquid filled ion chamber, and a range of small volume air filled ionization chambers (volumes ranging from 0.002 cm3 to 0.3 cm3). All detector measurements were corrected for volume averaging effect and compared with dose ratios determined from alanine to derive a detector correction factors that account for beam perturbation related to nonwater equivalence of the detector materials.
Results: For the detectors used in this study, volume averaging corrections ranged from unity for the smallest detectors such as the diodes, 1.148 for the 0.14 cm3 air filled ionization chamber and were as high as 1.924 for the 0.3 cm3 ionization chamber. After applying volume averaging corrections, the detector readings were consistent among themselves and with alanine measurements for several small detectors but they differed for larger detectors, in particular for some small ionization chambers with volumes larger than 0.1 cm3.
書誌情報 Medical Physics

巻 41, 号 7, p. 072103-1-072103-16, 発行日 2014-06
出版者
出版者 American Association of Physicists in Medicine
ISSN
収録物識別子タイプ ISSN
収録物識別子 0094-2405
DOI
識別子タイプ DOI
関連識別子 10.1118/1.4883795
関連サイト
識別子タイプ URI
関連識別子 http://scitation.aip.org/content/aapm/journal/medphys/41/7/10.1118/1.4883795
関連名称 http://scitation.aip.org/content/aapm/journal/medphys/41/7/10.1118/1.4883795
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 23:43:09.487104
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3