Patient-specific QA and delivery verification of scanned ion beam at NIRS-HIMAC

https://repo.qst.go.jp/records/46904
https://repo.qst.go.jp/records/46904
8939da11-b0a8-448a-aafd-169ed91a2706
Item type 学術雑誌論文 / Journal Article(1)
公開日 2014-11-06
タイトル
タイトル Patient-specific QA and delivery verification of scanned ion beam at NIRS-HIMAC
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Furukawa, Takuji

× Furukawa, Takuji

WEKO 468051

Furukawa, Takuji

Search repository
Inaniwa, Taku

× Inaniwa, Taku

WEKO 468052

Inaniwa, Taku

Search repository
Hara, Yousuke

× Hara, Yousuke

WEKO 468053

Hara, Yousuke

Search repository
Mizushima, Kota

× Mizushima, Kota

WEKO 468054

Mizushima, Kota

Search repository
Shirai, Toshiyuki

× Shirai, Toshiyuki

WEKO 468055

Shirai, Toshiyuki

Search repository
Noda, Kouji

× Noda, Kouji

WEKO 468056

Noda, Kouji

Search repository
古川 卓司

× 古川 卓司

WEKO 468057

en 古川 卓司

Search repository
稲庭 拓

× 稲庭 拓

WEKO 468058

en 稲庭 拓

Search repository
原 洋介

× 原 洋介

WEKO 468059

en 原 洋介

Search repository
水島 康太

× 水島 康太

WEKO 468060

en 水島 康太

Search repository
白井 敏之

× 白井 敏之

WEKO 468061

en 白井 敏之

Search repository
野田 耕司

× 野田 耕司

WEKO 468062

en 野田 耕司

Search repository
抄録
内容記述タイプ Abstract
内容記述 Purpose: To evaluate a patient-specific QA program and system for constancy checking of a scanning delivery system developed at the National Institute of Radiological Sciences.Methods: For the patient-specific QA, all the planned beams are recalculated on a water phantom with treatment planning software (TPS). The recalculated dose distributions are compared with the measured distributions using a 2D ionization chamber array at several depths, and evaluated using gamma index analysis with criteria of 3% and 3 mm and a pass rate of 90%. For the constancy check, the authors developed the multiwire proportional chamber (MWPC), which can record the delivered 2D fluence images in a slice-by-slice manner. During irradiation for dosimetric QA with the 2D ionization chamber array and an accordion-type water phantom, the 2D fluence images are recorded using the MWPC in the delivery system. These recorded images are then compared to those taken in the treatment session to check the constancy check. This analysis also employs gamma index analysis using the same criteria as in the patient-specific QA. These patient-specific QA and constancy check evaluations were performed using the data of 122 patients.Results: In the patient-specific QA, the measured dose distributions agreed well with those calculated by the TPS, and the QA criteria were satisfied in all measurements. The additional check of the fluence comparison ensured the constancy of the delivered field during each treatment irradiation.Conclusions: The authors established a patient-specific QA program and additional check of delivery constancy in every treatment session. Fluence comparison is a strong tool for constancy checking of the delivery system.
書誌情報 Medical Physics

巻 40, 号 12, p. 121707-1-121707-7, 発行日 2013-11
PubMed番号
識別子タイプ PMID
関連識別子 24320493
DOI
識別子タイプ DOI
関連識別子 doi:10.1118/1.4828845
関連サイト
識別子タイプ DOI
関連識別子 http://dx.doi.org/10.1118/1.4828845
関連名称 http://dx.doi.org/10.1118/1.4828845
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 23:46:13.855172
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3