WEKO3
アイテム
Intrinsic spatial resolution evaluation of the X'tal cube PET detector based on a 3D crystal block segmented by laser processing
https://repo.qst.go.jp/records/46490
https://repo.qst.go.jp/records/46490d9c1d9a6-0faa-4f3a-b387-c8980b888403
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2013-02-14 | |||||
タイトル | ||||||
タイトル | Intrinsic spatial resolution evaluation of the X'tal cube PET detector based on a 3D crystal block segmented by laser processing | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Yoshida, Eiji
× Yoshida, Eiji× Tashima, Hideaki× Inadama, Naoko× Nishikido, Fumihiko× Moriya, Takahiro× Omura, Tomohide× Watanabe, Mitsuo× Murayama, Hideo× Yamaya, Taiga× 吉田 英治× 田島 英朗× 稲玉 直子× 錦戸 文彦× 大村 知秀× 村山 秀雄× 山谷 泰賀 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | The X'tal cube is a depth-of-interaction (DOI)-PET detector which is aimed at obtaining isotropic resolution by effective readout of scintillation photons from the six sides of a crystal block. The X'tal cube is composed of the 3D crystal block with isotropic resolution and arrays of multi-pixel photon counters (MPPCs). In this study, to fabricate the 3D crystal block efficiently and precisely, we applied a sub-surface laser engraving (SSLE) technique to a monolithic crystal block instead of gluing segmented small crystals. The SSLE technique provided micro-crack walls which carve a groove into a monolithic scintillator block. Using the fabricated X'tal cube, we evaluated its intrinsic spatial resolution to show a proof of concept of isotropic resolution. The 3D grids of 2 mm pitch were fabricated into an 18 x 18 x 18 mm(3) monolithic lutetium yttrium orthosilicate (LYSO) crystal by the SSLE technique. 4 x 4 MPPCs were optically coupled to each surface of the crystal block. The X'tal cube was uniformly irradiated by (22)Na gamma rays, and all of the 3D grids on the 3D position histogram were separated clearly by an Anger-type calculation from the 96-channel MPPC signals. Response functions of the X'tal cube were measured by scanning with a (22)Na point source. The gamma-ray beam with a 1.0 mm slit was scanned in 0.25 mm steps by positioning of the X'tal cube at vertical and 45 incident angles. The average FWHM resolution at both incident angles was 2.1 mm. Therefore, we confirmed the isotropic spatial resolution performance of the X'tal cube. | |||||
書誌情報 |
Radiological Physics and Technology 巻 6, 号 1, p. 21-27, 発行日 2012-07 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 1865-0333 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1007/s12194-012-0167-0 |