Patient Handling System for Carbon-ion Beam Scanning Therapy
https://repo.qst.go.jp/records/46432
https://repo.qst.go.jp/records/4643252e57d3b-4a0c-48da-85d7-26cebd76fc1e
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2012-11-09 | |||||
タイトル | ||||||
タイトル | Patient Handling System for Carbon-ion Beam Scanning Therapy | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Mori, Shinichiro
× Mori, Shinichiro× Shirai, Toshiyuki× Takei, Yuka× Furukawa, Takuji× Inaniwa, Taku× Matsuzaki, Yuka× Kumagai, Motoki× Murakami, Takeshi× Noda, Kouji× 森 慎一郎× 白井 敏之× 武井 由佳× 古川 卓司× 稲庭 拓× 松崎 有華× 熊谷 始紀× 村上 健× 野田 耕司 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Our institution established a new treatment facility for carbon ion beam scanning therapy in 2010. The major advantages of scanning beam treatment compared to the passive beam treatment are the following: high dose conformation with less excessive dose to the normal tissues, no bolus compensator and patient collimator/multi-leaf collimator, better dose efficiency by reducing the number of scatters. The new facility was designed to solve several problems encountered in the existing facility, at which several thousand patients were treated over more than 15 years. Here, we introduce the patient handling system in the new treatment facility. The new facility incorporates three main systems, a scanning irradiation system (S-IR), treatment planning system (TPS), and patient handling system (PTH). The PTH covers a wide range of functions including imaging, geometrical/position accuracy including motion management (immobilization, robotic arm treatment bed), layout of the treatment room, treatment workflow, software, and others. The first clinical trials without respiratory gating have been successfully started. The PTH allows a reduction in patient stay in the treatment room to as few as 7 min. The PTH plays an important role in carbon ion beam scanning therapy at the new institution, particularly in the management of patient handling, application of image-guided therapy, and improvement of treatment workflow, and thereby allows substantially better treatment at minimum cost. | |||||
書誌情報 |
Journal of Applied Clinical Medical Physics (Online Only : http://www.jacmp.org/) 巻 13, 号 6, p. 226-240, 発行日 2012-11 |
|||||