System design of a small OpenPET prototype with 4-layer DOI detectors

https://repo.qst.go.jp/records/46285
https://repo.qst.go.jp/records/46285
01df4b39-b353-434b-acbc-b5f0ed4b6d1f
Item type 学術雑誌論文 / Journal Article(1)
公開日 2012-03-19
タイトル
タイトル System design of a small OpenPET prototype with 4-layer DOI detectors
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Yoshida, Eiji

× Yoshida, Eiji

WEKO 461038

Yoshida, Eiji

Search repository
Kinouchi, Shoko

× Kinouchi, Shoko

WEKO 461039

Kinouchi, Shoko

Search repository
Tashima, Hideaki

× Tashima, Hideaki

WEKO 461040

Tashima, Hideaki

Search repository
Nishikido, Fumihiko

× Nishikido, Fumihiko

WEKO 461041

Nishikido, Fumihiko

Search repository
Inadama, Naoko

× Inadama, Naoko

WEKO 461042

Inadama, Naoko

Search repository
Murayama, Hideo

× Murayama, Hideo

WEKO 461043

Murayama, Hideo

Search repository
Yamaya, Taiga

× Yamaya, Taiga

WEKO 461044

Yamaya, Taiga

Search repository
吉田 英治

× 吉田 英治

WEKO 461045

en 吉田 英治

Search repository
木内 尚子

× 木内 尚子

WEKO 461046

en 木内 尚子

Search repository
田島 英朗

× 田島 英朗

WEKO 461047

en 田島 英朗

Search repository
錦戸 文彦

× 錦戸 文彦

WEKO 461048

en 錦戸 文彦

Search repository
稲玉 直子

× 稲玉 直子

WEKO 461049

en 稲玉 直子

Search repository
村山 秀雄

× 村山 秀雄

WEKO 461050

en 村山 秀雄

Search repository
山谷 泰賀

× 山谷 泰賀

WEKO 461051

en 山谷 泰賀

Search repository
抄録
内容記述タイプ Abstract
内容記述 We have proposed an OpenPET geometry which consists of two axially separated detector rings. The open gap is suitable for in-beam PET. We have developed the small prototype of the OpenPET especially for a proof of concept of in-beam imaging. This paper presents an overview of the main features implemented in this prototype. We also evaluated the detector performance. This prototype was designed with 2 detector rings having 8 depth-of-interaction detectors. Each detector consisted of 784 Lu2xGd2(1-x)SiO5:Ce (LGSO) which were arranged in a 4-layer design, coupled to a position-sensitive photomultiplier tube (PS-PMT). The size of the LGSO array was smaller than the sensitive area of the PS-PMT, so that we could obtain sufficient LGSO identification. Peripheral LGSOs near the open gap directly detect the gamma rays on the side face in the OpenPET geometry. Output signals of two detectors stacked axially were projected onto one 2-dimensional position histogram for reduction of the scale of a coincidence processor. Front-end circuits were separated from the detector head by 1.2-m coaxial cables for the protection of electronic circuits from radiation damage. The detectors had sufficient crystal identification capability. Cross talk between the combined two detectors could be ignored. The timing and energy resolutions were 3.0 ns and 14%, respectively. The coincidence window was set 20 ns, because the timing histogram showed that not only the main peak, but also two small shifted peaks were caused by the coaxial cable. However, the detector offers the promise of sufficient performance, because random coincidences are at a nearly undetectable level for in-beam PET experiments.
書誌情報 Radiological Physics and Technology

巻 5, 号 1, p. 92-97, 発行日 2011-11
ISSN
収録物識別子タイプ ISSN
収録物識別子 1865-0333
DOI
識別子タイプ DOI
関連識別子 10.1007/s12194-011-0142-1
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-15 23:53:41.119486
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3