WEKO3
アイテム
A certified reference material for radionuclides in the water sample from Irish Sea (IAEA-443)
https://repo.qst.go.jp/records/46025
https://repo.qst.go.jp/records/46025eddb4767-84c1-428d-a126-b864c94039f0
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2011-04-04 | |||||
タイトル | ||||||
タイトル | A certified reference material for radionuclides in the water sample from Irish Sea (IAEA-443) | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
K, Pham M
× K, Pham M× Zheng, Jian× et.al× 鄭 建 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | A new Reference Material (RM) for radionuclides in sea water from the Irish Sea (IAEA-443) is described and the results of the certification process are presented. Ten radionuclides (3H, 40K, 90Sr, 137Cs, 234U, 235U, 238U, 238Pu, 239+240Pu and 241Am) have been certified, and information values on massic activities with 95% confidence intervals are given for four radionuclides (230Th, 232Th, 239Pu and 240Pu). Results for less frequently reported radionuclides (99Tc, 228Th, 237Np and 241Pu) are also reported. The RM can be used for quality assurance?quality control of the analysis of radionuclides in water samples, for the development and validation of analytical methods and for training purposes. The material is available in 5 liters units. | |||||
書誌情報 |
Journal of Radioanalytical and Nuclear Chemistry 巻 288, 号 2, p. 603-611, 発行日 2011-05 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0236-5731 |