Development of a Portable Graphite Calorimeter for Radiation Dosimetry

https://repo.qst.go.jp/records/45220
https://repo.qst.go.jp/records/45220
96a72c96-607d-4098-acb9-28ed7d286f87
Item type 学術雑誌論文 / Journal Article(1)
公開日 2008-07-08
タイトル
タイトル Development of a Portable Graphite Calorimeter for Radiation Dosimetry
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Sakama, Makoto

× Sakama, Makoto

WEKO 449146

Sakama, Makoto

Search repository
Kanai, Tatsuaki

× Kanai, Tatsuaki

WEKO 449147

Kanai, Tatsuaki

Search repository
Fukumura, Akifumi

× Fukumura, Akifumi

WEKO 449148

Fukumura, Akifumi

Search repository
坂間 誠

× 坂間 誠

WEKO 449149

en 坂間 誠

Search repository
金井 達明

× 金井 達明

WEKO 449150

en 金井 達明

Search repository
福村 明史

× 福村 明史

WEKO 449151

en 福村 明史

Search repository
抄録
内容記述タイプ Abstract
内容記述 Abstract
We developed and performance-tested a portable graphite calorimeter designed to measure the absolute
dosimetry of various beams including heavy-ion beams, using a flexible and convenient means of measurement.
This measurement system is fully remote-controlled by GPIB systems. The control system uses digital PID
(Proportional, Integral, Derivative) control methods based on a LabVIEW program1). It was possible to attain
stable conditions in a shorter time by this system. The standard deviation of the measurements using the
calorimeter was 0.79 % at a dose rate of 0.8 Gy/min in seventeen calorimeter runs of a 60Co photon beam. The
overall uncertainty for the absorbed dose to graphite and water of a 60Co photon beam using the developed
calorimeter was 0.89 % and 1.35 %, respectively. Estimations of the correction factors due to vacuum gaps,
impurities in the core, the dose gradient and the radiation profile were included in the uncertainty. The
absorbed doses to graphite and water irradiated by a 60Co photon beam were compared with dosimetry using
ionization chambers. The absorbed doses to graphite and water estimated by the two dosimetry methods agreed
within 0.1 % and 0.3 %, respectively.
書誌情報 Japanese Journal of Medical Physics

巻 28, 号 1, p. 1-14, 発行日 2008-05
ISSN
収録物識別子タイプ ISSN
収録物識別子 1345-5354
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-16 00:06:19.138732
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3