Optimization for fast-scanning irradiation in particle therapy

https://repo.qst.go.jp/records/44971
https://repo.qst.go.jp/records/44971
4090e9e8-c8af-4042-aca5-f64171a603a0
Item type 学術雑誌論文 / Journal Article(1)
公開日 2007-09-10
タイトル
タイトル Optimization for fast-scanning irradiation in particle therapy
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Inaniwa, Taku

× Inaniwa, Taku

WEKO 446573

Inaniwa, Taku

Search repository
Furukawa, Takuji

× Furukawa, Takuji

WEKO 446574

Furukawa, Takuji

Search repository
Tomitani, Takehiro

× Tomitani, Takehiro

WEKO 446575

Tomitani, Takehiro

Search repository
Satou, Shinji

× Satou, Shinji

WEKO 446576

Satou, Shinji

Search repository
Noda, Kouji

× Noda, Kouji

WEKO 446577

Noda, Kouji

Search repository
Kanai, Tatsuaki

× Kanai, Tatsuaki

WEKO 446578

Kanai, Tatsuaki

Search repository
稲庭 拓

× 稲庭 拓

WEKO 446579

en 稲庭 拓

Search repository
古川 卓司

× 古川 卓司

WEKO 446580

en 古川 卓司

Search repository
富谷 武浩

× 富谷 武浩

WEKO 446581

en 富谷 武浩

Search repository
佐藤 眞二

× 佐藤 眞二

WEKO 446582

en 佐藤 眞二

Search repository
野田 耕司

× 野田 耕司

WEKO 446583

en 野田 耕司

Search repository
金井 達明

× 金井 達明

WEKO 446584

en 金井 達明

Search repository
抄録
内容記述タイプ Abstract
内容記述 In three-dimensional irradiation with pencil beam scanning, an extra dose is inevitably delivered to the irradiated site due to the finite reaction times of the beam delivery system, and it causes a severe distortion of the dose distribution in the target region. Since the amount of the extra dose is proportional to the beam intensity, the dose uniformity deteriorates as the beam intensity is increased in order to shorten the treatment time. In order to overcome this problem and shorten the treatment time, we have developed an optimization method in which the extra dose is integrated into the optimization process of the best weighting matrix. The effectiveness and applicability of the optimization method for spot and raster scanning irradiation were confirmed with computer simulations and also confirmed using irradiation experiments for spot scanning irradiation. The treatment time could be shortened to about one sixth of the time needed without taking the extra dose into account while obtaining the same degree of dose homogeneity in the target volume. A typical treatment time with the proposed method is about 15 s for the irradiation of a spherical target with an 80 mm diameter at 3 GyE.
書誌情報 Medical Physics

巻 34, 号 8, p. 3302-3311, 発行日 2007-07
ISSN
収録物識別子タイプ ISSN
収録物識別子 0094-2405
DOI
識別子タイプ DOI
関連識別子 10.1118/1.2712572
戻る
0
views
See details
Views

Versions