  1. 原著論文

Production design and evaluation of a novel breast phantom with various breast glandular fractions

https://repo.qst.go.jp/records/44953
https://repo.qst.go.jp/records/44953
8a957365-2406-44dc-b0c9-586ce34f6df4
Item type 学術雑誌論文 / Journal Article(1)
公開日 2007-08-24
タイトル
タイトル Production design and evaluation of a novel breast phantom with various breast glandular fractions
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Fujisaki, Tatsuya

× Fujisaki, Tatsuya

WEKO 446426

Fujisaki, Tatsuya

Search repository
Kimura, Masaya

× Kimura, Masaya

WEKO 446427

Kimura, Masaya

Search repository
Saitoh, Hidetoshi

× Saitoh, Hidetoshi

WEKO 446428

Saitoh, Hidetoshi

Search repository
Abe, Shinji

× Abe, Shinji

WEKO 446429

Abe, Shinji

Search repository
Hiraoka, Takeshi

× Hiraoka, Takeshi

WEKO 446430

Hiraoka, Takeshi

Search repository
阿部 慎司

× 阿部 慎司

WEKO 446431

en 阿部 慎司

Search repository
平岡 武

× 平岡 武

WEKO 446432

en 平岡 武

Search repository
抄録
内容記述タイプ Abstract
内容記述 The purpose of this study was to evaluate the production design of a novel breast phantom, which has adjustable breast glandular fractions and potential application in the mammography quality assurance/quality control system. The breast phantom was based on a urethane resin that was used to adjust the breast glandular fraction by varying the amount of plasticizer added. The resin was cured at constant temperature and humidity. Theoretical phantom properties, such as elemental composition, specific density, effective atomic number, electron density, and linear attenuation coefficients, at various energies were compared to those of breast tissue tabulated in the ICRU 44. These properties were also compared to polymethyl methacrylate resin and BR12. The novel breast phantom was made to represent breast glandular content calculated from breast tissue of the ICRU 44. We hypothesized that the breast phantom theoretical properties are approximately equal to those of the BR12, with is known for being an excellent substitute breast phantom. It was found that the phantom can be used to improve both mammography performance and dosimetry.
書誌情報 Radiation Medicine

巻 24, 号 10, p. 647-652, 発行日 2006-12
ISSN
収録物識別子タイプ ISSN
収録物識別子 0288-2043
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-16 00:09:24.184259
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3