WEKO3
アイテム
New shielding materials for clinical electron beams
https://repo.qst.go.jp/records/44621
https://repo.qst.go.jp/records/44621e9eeb933-0cfc-4214-9f1b-52fb68e88f05
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2006-09-25 | |||||
タイトル | ||||||
タイトル | New shielding materials for clinical electron beams | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Tajiri, Minoru
× Tajiri, Minoru× Tokiya, Yuuji× Uenishi, Jun× Sunaoka, Masayoshi× Watanabe, Kazuhiro× 田尻 稔× 鴇矢 祐治× 上西 潤× 砂岡 正良× 渡邊 和洋 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Since lead has recently been recognized as a source of environmental pollution, we have investigated new electron shielding materials that do not contain lead. We compared the shielding thicknesses of a hard plate and a sheet composed of the new materials with that of lead for electron beams. The shielding thickness was evaluated as the thickness required for shielding primary electrons. The comparison revealed the shielding ability of the hard plate and sheet is approximately equivalent to 1.0 and 0.9 times that of lead, respectively. The thickness (in millimeters) required for shielding by the hard-plate, as well as the thickness of lead, is related to approximately half of the electron-beam energy (in MeV). The shielding ability of the sheet is also equivalent to that of Lipowitz alloy. Moreover these materials are environmentally friendly, and can be easily customized into arbitrary shapes. Therefore they can be used as lead substitutes for shielding against electron beams. | |||||
書誌情報 |
Radiotherapy and Oncology 巻 80, 号 3, p. 391-393, 発行日 2006 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0167-8140 |