Development of a 3D CT-Scanner Using a Cone Beam and Video-Fluoroscopic System

https://repo.qst.go.jp/records/44373
https://repo.qst.go.jp/records/44373
27803cfa-6ad4-46c8-9961-9667b2c31ada
Item type 学術雑誌論文 / Journal Article(1)
公開日 2006-06-29
タイトル
タイトル Development of a 3D CT-Scanner Using a Cone Beam and Video-Fluoroscopic System
言語
言語 jpn
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 遠藤, 真広

× 遠藤, 真広

WEKO 440922

遠藤, 真広

Search repository
吉田, 勝哉

× 吉田, 勝哉

WEKO 440923

吉田, 勝哉

Search repository
日下部, 正宏

× 日下部, 正宏

WEKO 440924

日下部, 正宏

Search repository
舘野, 之男

× 舘野, 之男

WEKO 440925

舘野, 之男

Search repository
その他

× その他

WEKO 440926

その他

Search repository
遠藤 真広

× 遠藤 真広

WEKO 440927

en 遠藤 真広

Search repository
吉田 勝哉

× 吉田 勝哉

WEKO 440928

en 吉田 勝哉

Search repository
日下部 正宏

× 日下部 正宏

WEKO 440929

en 日下部 正宏

Search repository
舘野 之男

× 舘野 之男

WEKO 440930

en 舘野 之男

Search repository
抄録
内容記述タイプ Abstract
内容記述 We describe the design and implementation of a system that acquires three-dimensional (3D) data of high-contrast objects such as bone, lung, and blood vessels (enhanced by contrast agent). This 3D computed tomography (CT) system is based on a cone beam and video-fluoroscopic system and yields data that is amenable to 3D image processing. An X-ray tube and a large area two-dimensional detector were mounted on a single frame and rotated around objects in 12 seconds. The large area detector consisted of a fluorescent plate and a charge coupled device (CCD) video camera. While tha X-ray tube was rotated around the object, a pulsed X-ray was generated (30 pulses per second) and 360 projected images were collected in a 12-second scan. A 256x256x256 matrix image was reconstructed using a high-speed parallel processor. Reconstruction required approximately 6 minutes. Two volunteers under-went scans of the head or chest. High-contrast objects such as bronchial, vascular, and mediastinal structures in the thorax, or bones ans air cavities in the head were delineated in a "real" 3D format. Our 3D CT-scanner appears to produce data useful for clinical imaging and 3D image processing.
書誌情報 Radiation Medicine

巻 16, 号 1, p. 7-12, 発行日 1998
ISSN
収録物識別子タイプ ISSN
収録物識別子 0288-2043
戻る
0
views
See details
Views

Versions

Ver.1 2023-05-16 00:16:12.976645
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3