  1. 原著論文

Detection of Small Degree of Nonuniformity in Dialysate Flow in Hollow-Fiber Dialyzer Using Proton Magnetic Resonace Imaging

https://repo.qst.go.jp/records/43773
https://repo.qst.go.jp/records/43773
76b4e02d-59fa-48dc-a02d-7747daeae4ac
Item type 学術雑誌論文 / Journal Article(1)
公開日 2005-08-17
タイトル
タイトル Detection of Small Degree of Nonuniformity in Dialysate Flow in Hollow-Fiber Dialyzer Using Proton Magnetic Resonace Imaging
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
アクセス権
アクセス権 metadata only access
アクセス権URI http://purl.org/coar/access_right/c_14cb
著者 Osuga, Toshiaki

× Osuga, Toshiaki

WEKO 435235

Osuga, Toshiaki

Search repository
Obata, Takayuki

× Obata, Takayuki

WEKO 435236

Obata, Takayuki

Search repository
Ikehira, Hiroo

× Ikehira, Hiroo

WEKO 435237

Ikehira, Hiroo

Search repository
大須賀 敏明

× 大須賀 敏明

WEKO 435238

en 大須賀 敏明

Search repository
小畠 隆行

× 小畠 隆行

WEKO 435239

en 小畠 隆行

Search repository
池平 博夫

× 池平 博夫

WEKO 435240

en 池平 博夫

Search repository
抄録
内容記述タイプ Abstract
内容記述 A small degree of nonuniformity in dialysate flow in a hollow-fiber dialyzer was detected using proton magnetic resonance imaging(MRI). Since paramagnetic ions reduce the spin-lattice relaxation time of protons around them, MRI can detect Gd in water. An aqueous solution of a chelate compound of Gd was impulsively injected into the dialysate flow path at a flow rate 500 cm3/m, which is that utilized in actual dialysis. Despite the apparent elimination of Gd from the dialysate flow path by the newly injected dialysate fluid after the injection of Gd was terminated, MRI revealed that Gd remained in th interior of the hollow fiber. The observed structure pattern of the Gd concentration profile revealed that the dialysate flow had a small degree of nununiformity despite the currently established design to restrict channeling in dialysate flow. Local nonuniformity of the hollow-fiber density and vortex generation in the dialysate flow were considered to cause the nonuniformity in the dialysate flow.
書誌情報 Magnetic Resonance Imaging

巻 22, 号 3, p. 417-420, 発行日 2004-04
ISSN
収録物識別子タイプ ISSN
収録物識別子 0730-725X
戻る
0
views
See details
Views