WEKO3
アイテム
Image reconstruction from limited angle Compton camera deta
https://repo.qst.go.jp/records/43380
https://repo.qst.go.jp/records/43380086136a7-385b-42c1-9ea4-fb5fcbe2d7db
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2004-05-05 | |||||
タイトル | ||||||
タイトル | Image reconstruction from limited angle Compton camera deta | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Tomitani, Takehiro
× Tomitani, Takehiro× Hirasawa, Masahiko× 富谷 武浩× 平澤 雅彦 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | The Compton camera is used for imaging the distributions of gamma-ray direction in gamma-ray telescope for astrophysics and for imaging radioisotope distributions in nuclear medicine without the need for collimators. The integration of gamma-rays on a cone is measured with the camera, so that some sort of inversion method is needed. Parra found an analytical inversion algorithm based on spherical harmonics expansion of projection data. His algorithm is applicable to the full set of projection data. In this paper, six possible reconstruction algorithms that allow image reconstruction from projections with a finite range of scattering angles are investigated. Four algorithms have instability problems and two other are practical. However, the variance of the reconstructed image diverges in these two cases, so that window functions are introduced with which the variance becomes finite at a cost of spatial resolution. These two algorithms are compared in terms of variance. The algorithm based on the inversion of the summed back-projection is superior to the algorithm based on the inversion of the summed projection. |
|||||
書誌情報 |
Physics in Medicine and Biology 巻 47, p. 2129-2146, 発行日 2002-06 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0031-9155 |