| アイテムタイプ |
会議発表用資料 / Presentation(1) |
| 公開日 |
2026-06-22 |
| タイトル |
|
|
タイトル |
Calibration of a dose calibrator using in-house produced Ac-225 |
|
言語 |
en |
| 言語 |
|
|
言語 |
eng |
| 資源タイプ |
|
|
資源タイプ識別子 |
http://purl.org/coar/resource_type/c_6670 |
|
資源タイプ |
conference poster |
| 著者 |
Ichinose Jun
Yasushi Sato
Chihiro Shimodan
Nagatsu Kotaro
|
| 抄録 |
|
|
内容記述 |
Actinium-225 (Ac-225) is a highly promising alpha emitter for targeted radionuclide therapy (TRT). Activity is a critical parameter not only for product quality but also for process control during production. This study evaluates the time-dependent calibration of a commercial dose calibrator prior to the establishment of radioactive equilibrium. Ac-225 produced via proton irradiation of radium-226 (Ra226) was measured, and the absolute activity was determined using the Triple-to-Double Coincidence Ratio (TDCR) method. A time-dependent calibration factor was derived, enabling accurate activity determination under pre-equilibrium conditions. |
| 会議概要(会議名, 開催地, 会期, 主催者等) |
|
|
内容記述 |
WTTC20 (20th Workshop on Targetry and Target Chemistry) |
| 発表年月日 |
|
|
日付 |
2026-08-23 |