Sub-second dynamic rat imaging with an ultrasensitive small animal PET scanner

https://repo.qst.go.jp/records/2002307
https://repo.qst.go.jp/records/2002307
c42cea4a-85a4-4d9e-b35f-d5619241b33f
アイテムタイプ 会議発表用資料 / Presentation(1)
公開日 2025-08-07
タイトル
タイトル Sub-second dynamic rat imaging with an ultrasensitive small animal PET scanner
言語 en
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_c94f
資源タイプ conference presentation
著者 Hangyu Kang

× Hangyu Kang

Hangyu Kang

Search repository
Tashima Hideaki

× Tashima Hideaki

Tashima Hideaki

Search repository
Wakizaka Hidekatsu

× Wakizaka Hidekatsu

Wakizaka Hidekatsu

Search repository
Akamatsu Go

× Akamatsu Go

Akamatsu Go

Search repository
Iwao Yuma

× Iwao Yuma

Iwao Yuma

Search repository
Toramatsu Chie

× Toramatsu Chie

Toramatsu Chie

Search repository
Yamaya Taiga

× Yamaya Taiga

Yamaya Taiga

Search repository
抄録
内容記述 Dynamic positron emision tomography (PET) imaging can provide molecular information of living objects as a function of time. However, the temporal resolution of the state-of-the art PET scanners is over 1 second, which is not fast enough to capture rapidely changing brain or cardiac functions. Previously, we developed the total-body small animal (TBS) PET scanner with 4-layer depth-of-interaction (DOI) detectors to achieve ultrahigh sensitivity. However, the physical performance was not fully evaluated. More importantly, in vivo dynamic imaging was not demonstrated. In this study, we present the physical performance and sub-second dynamic PET imaging results of a rat with the TBS-PET scanner. The TBS PET has 155 mm inner diameter and 325.6 mm axial coverage featuring 45% peak sensitivity with an energy window of 250-750 keV (Fig. 1). Each DOI PET detector consisted of 4-layer 16×16 GSOZ crystals (2.8×2.8×7.5 mm3) and an 8×8 array multi-anode PMT. A bolus of 18F-FDG (21.2 MBq) was injected into an anesthetized rat via tail vein followed by saline injection. Dynamic PET images of a rat were obtained with 0.5 s time frame. In the time activity curves, the time dispersion between the left and right ventricles (1 s) was clearly separated. In conclusion, the TBS-PET enabled sub-second dynamic imaging for rodent models. The detailed dynamic PET analysis results will be presented at the conference.
会議概要(会議名, 開催地, 会期, 主催者等)
内容記述 The 86th JSAP Autumn Meeting 2025
発表年月日
日付 2025-09-09
戻る
0
views
See details
Views

Versions

Ver.1 2026-01-19 00:20:15.497860
Show All versions

Share

Share
tweet

Cite as

Other

print

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX
  • ZIP

コミュニティ

確認

確認

確認


Powered by WEKO3


Powered by WEKO3