ログイン
言語:

WEKO3

  • トップ
  • ランキング
To
lat lon distance
To

Field does not validate



インデックスリンク

インデックスツリー

メールアドレスを入力してください。

WEKO

One fine body…

WEKO

One fine body…

アイテム

  1. 原著論文

Energy resolution improvement in simplified alpha spectroscopy for radionuclidic purity tests on 225Ac production floors

https://repo.qst.go.jp/records/2001316
https://repo.qst.go.jp/records/2001316
0188ed57-d659-43ed-926a-e5fc69b94951
Item type 学術雑誌論文 / Journal Article(1)
公開日 2024-09-25
タイトル
タイトル Energy resolution improvement in simplified alpha spectroscopy for radionuclidic purity tests on 225Ac production floors
言語 en
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Mihei Yuma

× Mihei Yuma

Mihei Yuma

Search repository
Tani Kotaro

× Tani Kotaro

Tani Kotaro

Search repository
Ichinose Jun

× Ichinose Jun

Ichinose Jun

Search repository
Nagatsu Kotaro

× Nagatsu Kotaro

Nagatsu Kotaro

Search repository
Fukushi Masahiro

× Fukushi Masahiro

Fukushi Masahiro

Search repository
Kurihara Osamu

× Kurihara Osamu

Kurihara Osamu

Search repository
Inoue Kazumasa

× Inoue Kazumasa

Inoue Kazumasa

Search repository
抄録
内容記述タイプ Abstract
内容記述 This study investigated the feasibility of a simplified method of alpha spectroscopy for radionuclidic purity tests at 225Ac production sites that eliminates the need for a vacuum chamber. The impact of enhancing the energy resolution using a collimator was evaluated through radiation transport simulations. The results showed that a full width at tenth maximum (FWTM) of less than 300 keV was achieved for alpha particles from 241Am, for which the main energy peak was 5.5 MeV. Experimental validation using an electrodeposition source containing 237Np, 241Am, and 244Cm confirmed an FWTM of 272 keV for both 241Am and 244Cm. These two peaks, with a difference of approximately 300 keV, were effectively separated. In response to the growing demand for targeted radioisotope therapy, this simplified alpha spectroscopy method offers the potential to detect 226Ra mixed with 225Ac generated by accelerators, given the alpha energy difference of about 700 keV.
書誌情報 Radiation Protection Dosimetry

巻 200, 号 15, p. 1443-1449, 発行日 2024-09
出版者
出版者 Oxford University Press
DOI
識別子タイプ DOI
関連識別子 10.1093 / rpd / ncae185
戻る
0
views
See details
Views

Versions

Ver.1 2025-08-15 05:48:55.210799
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3