  1. 原著論文

Optimization of a YAP(Ce) prompt X-ray camera for imaging with spot scanning proton beams at clinical dose levels

https://repo.qst.go.jp/records/2000911
https://repo.qst.go.jp/records/2000911
33d117e9-6452-4ece-a771-2e2eed9cd5d7
アイテムタイプ 学術雑誌論文 / Journal Article(1)
公開日 2025-04-28
タイトル
タイトル Optimization of a YAP(Ce) prompt X-ray camera for imaging with spot scanning proton beams at clinical dose levels
言語 en
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Yamamoto Seiichi (Waseda Univ.)

× Yamamoto Seiichi (Waseda Univ.)

Yamamoto Seiichi (Waseda Univ.)

Search repository
Yamashita Tomohiro (Kobe Proton Center)

× Yamashita Tomohiro (Kobe Proton Center)

Yamashita Tomohiro (Kobe Proton Center)

Search repository
Kobayashi Yusuke (Kobe Proton Center)

× Kobayashi Yusuke (Kobe Proton Center)

Kobayashi Yusuke (Kobe Proton Center)

Search repository
Yabe Takuya

× Yabe Takuya

Yabe Takuya

Search repository
Akagi Takashi (Hyogo Ion Beam Medical Center)

× Akagi Takashi (Hyogo Ion Beam Medical Center)

Akagi Takashi (Hyogo Ion Beam Medical Center)

Search repository
Yamaguchi Mitsutaka

× Yamaguchi Mitsutaka

Yamaguchi Mitsutaka

Search repository
Kawachi Naoki

× Kawachi Naoki

Kawachi Naoki

Search repository
Kamada Kei (Tohoku Univ.)

× Kamada Kei (Tohoku Univ.)

Kamada Kei (Tohoku Univ.)

Search repository
Yoshikawa Akira (Tohoku Univ.)

× Yoshikawa Akira (Tohoku Univ.)

Yoshikawa Akira (Tohoku Univ.)

Search repository
Kataoka Jun (Waseda Univ.)

× Kataoka Jun (Waseda Univ.)

Kataoka Jun (Waseda Univ.)

Search repository
抄録
内容記述タイプ Abstract
内容記述 Prompt secondary electron bremsstrahlung X-ray (prompt X-ray) imaging using a low-energy X-ray camera is a promising method for observing the beam shape from outside a subject. However, it has sometimes been necessary to conduct such imaging at a higher dose than the clinical level to acquire images with acceptable quality. To solve this problem, we optimized a prompt X-ray imaging system to use for spot scanning proton therapy system. The new camera had more than one order higher sensitivity to image several types of beams, including those at the clinical dose level. The optimized prompt X-ray imaging system uses a 4 mm diameter pinhole collimator to increase sensitivity, and it is combined with a larger YAP(Ce) scintillator to increase the magnification ratio and thus improve spatial resolution. We used a list-mode data-acquisition system with high count rate capability. Prompt X-ray images were acquired by irradiating a water phantom with proton beams from the spot scanning proton therapy system. Measurements were taken for pencil beams, spread-out Bragg peak (SOBP) beams, and a beam utilized in actual clinical therapy. For all of the beams, we could measure scanning spot images within a spill and evaluate the ranges for the accumulated images at the clinical dose level. From the list-mode data, we measured the temporarily altered positions of the scanning beams as well as the accumulations of the prompt X-ray images. The optimized prompt X-ray imaging system could improve sensitivity while maintaining better spatial resolution. The new system realized prompt X-ray imaging at the clinical dose level and holds promise for future clinical imaging of prompt X-rays.
書誌情報 Journal of Instrumentation (JINST)

巻 18, p. T12005, 発行日 2023-04
出版者
出版者 IOP Publishing
ISSN
収録物識別子タイプ ISSN
収録物識別子 1748-0221
DOI
識別子タイプ DOI
関連識別子 10.1088/1748-0221/18/12/T12005
戻る
0
views
See details
Views

Versions

Ver.1 2025-08-15 03:05:03.601418
Show All versions