ログイン
Language:

WEKO3

  • トップ
  • ランキング
To
lat lon distance
To

Field does not validate



インデックスリンク

インデックスツリー

メールアドレスを入力してください。

WEKO

One fine body…

WEKO

One fine body…

アイテム

  1. 原著論文

Evaluation of Patient-Specific Quality Assurance for Carbon Ion Radiotherapy Using Full Energy Scanning Method at QST Hospital

https://repo.qst.go.jp/records/2000513
https://repo.qst.go.jp/records/2000513
b4f6cf45-d5ad-4551-a5b5-f9573a40e0c9
アイテムタイプ 学術雑誌論文 / Journal Article(1)
公開日 2024-06-07
タイトル
タイトル Evaluation of Patient-Specific Quality Assurance for Carbon Ion Radiotherapy Using Full Energy Scanning Method at QST Hospital
言語 en
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Inomata Asami

× Inomata Asami

Inomata Asami

Search repository
Katsumata Masashi

× Katsumata Masashi

Katsumata Masashi

Search repository
Sung-Hyun Lee

× Sung-Hyun Lee

Sung-Hyun Lee

Search repository
Suzuki Yui

× Suzuki Yui

Suzuki Yui

Search repository
Takeo Nakajima

× Takeo Nakajima

Takeo Nakajima

Search repository
Furuichi Wataru

× Furuichi Wataru

Furuichi Wataru

Search repository
Yamaoka Keisi

× Yamaoka Keisi

Yamaoka Keisi

Search repository
Yamamoto Atsushi

× Yamamoto Atsushi

Yamamoto Atsushi

Search repository
Mizuno Hideyuki

× Mizuno Hideyuki

Mizuno Hideyuki

Search repository
Kohno Ryosuke

× Kohno Ryosuke

Kohno Ryosuke

Search repository
抄録
内容記述タイプ Abstract
内容記述 Purpose: Patient-specific QA (PSQA) measurements for carbon ion radiotherapy (CIRT) are critical components of processes designed to identify
discrepancies between calculated and delivered doses. We report the results
of PSQA conducted at the QST Hospital during the period from September
2017 to March 2018. Methods: We analyzed PSQA results for 1448 fields
for 10 disease sites with various target volumes, target depths and number
of energy layers. For the PSQA, all the planned beams were recalculated on
a water phantom with treatment planning software. The recalculated dose
distributions were compared with the measured distributions using a 2D ionization chamber array at three depths, including 95% of the area of the prescription dose. These recalculated dose distributions were evaluated using
the 3%/3mm gamma index with a passing threshold of 90%. Results: The
passing rates for prostate, head and neck, and bone and soft tissue were 96.8%,
99.3%, and 91.7%, respectively. Additionally, 94.7% of lung plans with low
energy beams passed. Overall, the CIRT in the QST Hospital reached a high
passing rate of more than 95%. Although the remaining 5% failed to pass,
there was no dependence between measurement depth and disease sites in
these failures. Conclusion: Using PSQA measurements, we confirmed consistency between the planned and delivered doses for CIRT using the full
energy scanning method.
書誌情報 International Journal of Medical Physics, Clinical Engineering and Radiation Oncology

巻 11, 号 4, p. 200-209, 発行日 2022-11
出版者
出版者 Scientific Research Publishing
ISSN
収録物識別子タイプ ISSN
収録物識別子 2168-5444
DOI
識別子タイプ DOI
関連識別子 10.4236/ijmpcero.2022.114017
戻る
0
views
See details
Views

Versions

Ver.1 2025-08-15 02:22:11.973753
Show All versions

Share

Share
tweet

Cite as

Other

print

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX
  • ZIP

コミュニティ

確認

確認

確認


Powered by WEKO3


Powered by WEKO3