  1. 原著論文

CONTRIBUTION OF ORGAN-BASED TUBE CURRENT MODULATION TO THE REDUCTION OF LENS EXPOSURE DOSE IN HEAD 4D CT IMAGING: A PHANTOM STUDY.

https://repo.qst.go.jp/records/2000429
https://repo.qst.go.jp/records/2000429
1bac6fc6-3058-4a04-a1ad-518837a3fe79
アイテムタイプ 学術雑誌論文 / Journal Article(1)
公開日 2024-05-02
タイトル
タイトル CONTRIBUTION OF ORGAN-BASED TUBE CURRENT MODULATION TO THE REDUCTION OF LENS EXPOSURE DOSE IN HEAD 4D CT IMAGING: A PHANTOM STUDY.
言語 en
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Keisuke Nagamoto

× Keisuke Nagamoto

Keisuke Nagamoto

Search repository
Ryo Watanabe

× Ryo Watanabe

Ryo Watanabe

Search repository
Tomonori Kawachino

× Tomonori Kawachino

Tomonori Kawachino

Search repository
Yoshitaka Ohishi

× Yoshitaka Ohishi

Yoshitaka Ohishi

Search repository
Aina Yamamoto

× Aina Yamamoto

Aina Yamamoto

Search repository
Takashi Moritake

× Takashi Moritake

Takashi Moritake

Search repository
Shun-Ichi Nihei

× Shun-Ichi Nihei

Shun-Ichi Nihei

Search repository
Masayuki Kamochi

× Masayuki Kamochi

Masayuki Kamochi

Search repository
Naoki Kunugita

× Naoki Kunugita

Naoki Kunugita

Search repository
抄録
内容記述タイプ Abstract
内容記述 The purpose of this study was to evaluate the effectiveness of organ effect modulation (OEM) in reducing the lens dose in 4D computed tomography (CT) of the head in volume-acquisition (NVA) mode. Six radiophotoluminescent dosemeters were placed on the head of a RANDO phantom. The doses absorbed by the organs and image noise change rate were determined. The lens doses without OEM (i.e. in the OEMoff case) were higher than those with the same target standard deviation and volume-computed tomography dose index (CTDIvol) as in the OEMoff case (p?<?0.01). The image noise change rate was 11%. OEM reduced the lens dose during head 4D CT imaging in the NVA mode by 18%. Furthermore, the feasibility of lens dose reduction while ensuring sufficient image quality was confirmed under the condition in which OEM was employed with the same CTDIvol as in the OEMoff case.
書誌情報 Radiation protection dosimetry

巻 198, 号 18, p. 1377-1386, 発行日 2022-09
出版者
出版者 Oxford University Press
ISSN
収録物識別子タイプ ISSN
収録物識別子 1742-3406
PubMed番号
識別子タイプ PMID
関連識別子 36062449
DOI
識別子タイプ DOI
関連識別子 10.1093/rpd/ncac179
戻る
0
views
See details
Views

Versions

Ver.1 2025-08-15 02:19:27.546046