  1. 原著論文

Development and clinical verification of a small intestine motility measurement system using an ileus tube

https://repo.qst.go.jp/records/2000307
https://repo.qst.go.jp/records/2000307
430d0efc-b86a-4b16-93dd-dd2025d3dbcd
名前 / ファイル ライセンス アクション
cf63b14a83ef689e6fb84610b1a7f073.pdf Development and Clinical Verification of a Small Intestine Motility Measurement System Using an Ileus Tube.pdf (4.3 MB)
アイテムタイプ 学術雑誌論文 / Journal Article(1)
公開日 2023-09-08
タイトル
タイトル Development and clinical verification of a small intestine motility measurement system using an ileus tube
言語 en
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Shinji HOSOKAWA

× Shinji HOSOKAWA

Shinji HOSOKAWA

Search repository
Akihiro NAGANAWA

× Akihiro NAGANAWA

Akihiro NAGANAWA

Search repository
Takeshi SEKI

× Takeshi SEKI

Takeshi SEKI

Search repository
Kiyoshi Oka

× Kiyoshi Oka

Kiyoshi Oka

Search repository
Noriaki MANABE

× Noriaki MANABE

Noriaki MANABE

Search repository
Ken HARUMA

× Ken HARUMA

Ken HARUMA

Search repository
Junji YOSHINO

× Junji YOSHINO

Junji YOSHINO

Search repository
抄録
内容記述タイプ Abstract
内容記述 Intestinal obstruction is generally treated with an ileus tube. Success of treating functional ileus due to abnormal intestinal motility is judged through auscultation and medical interviews. Therefore, functional tests to visually evaluate the motility of the small intestine are important to elucidate the pathophysiology and observe the recovery process. An internal pressure measurement method that directly measures small intestine motility has been explored as a type of functional test. However, because the catheter is not in contact with the inner intestine wall, it is possible to measure the peristalsis movement, which is a large motility, but not the automatic rhythmic movement, which is a small motility. Therefore, we have developed a small intestine motility measurement system using an ileus tube. Small intestine motility is measured using the pressure applied to a balloon attached to an ileus tube. Peristaltic and automatic movement can be measured as the balloon contacts the inner intestinal wall, allowing more detailed evaluation of small intestinal motility. This study aimed at determining whether this measurement system can be applied clinically to accurately measure the small intestine motility. Hence, we first applied pressure to the balloon of the ileus tube using a pressurizing device to evaluate the basic performance ex vivo. We then conducted preliminary tests on a healthy volunteer. Analysis of the pressure test results showed that although the time delay of the balloon pressure measurement with respect to pressurization status was 4?6%, the cycles of the balloon pressure and displacement were consistent. Thus, it was concluded that the time delay had no effect on the proposed small intestine motility measurement system. In the preliminary test in a healthy volunteer, the pressure waveforms obtained during testing revealed a mixture of large and small pressure values of ~2 kPa at ~1?2 cycles/min and ~0.5 kPa at ~6?11 cycles/min, respectively. We confirmed that the larger pressure was similar to a peristalsis motility pattern, whereas the smaller pressure was similar to an automatic motility pattern. These results demonstrate that our measurement system is able to measure the small intestine motility and thus can be applied to comprehensive and visual measurement of small intestine motility.
書誌情報 Advanced Biomedical Engineering

巻 11, p. 228-236, 発行日 2022-11
出版者
出版者 Japanese Society for Medical and Biological Engineering
ISSN
収録物識別子タイプ ISSN
収録物識別子 2187-5219
DOI
識別子タイプ DOI
関連識別子 10.14326/abe.11.228
戻る
0
views
See details
Views

Versions

Ver.1 2025-08-15 01:53:22.956537
Show All versions

Share

Share
tweet

Cite as

Other

print

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX
  • ZIP

コミュニティ

確認

確認

確認


Powered by WEKO3


Powered by WEKO3