  1. 原著論文

DOSE MEASUREMENT PRECISION OF AN RPLD-BASED EYE LENS DOSEMETER APPLICABLE TO THE MEDICAL SECTOR

https://repo.qst.go.jp/records/2000165
https://repo.qst.go.jp/records/2000165
57ace444-c647-421b-8e51-f2d81860f225
アイテムタイプ 学術雑誌論文 / Journal Article(1)
公開日 2023-06-28
タイトル
タイトル DOSE MEASUREMENT PRECISION OF AN RPLD-BASED EYE LENS DOSEMETER APPLICABLE TO THE MEDICAL SECTOR
言語 en
言語
言語 eng
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ journal article
著者 Kowatari Munehiko

× Kowatari Munehiko

Kowatari Munehiko

Search repository
keisuke Nagamoto

× keisuke Nagamoto

keisuke Nagamoto

Search repository
Nakagami Koichi

× Nakagami Koichi

Nakagami Koichi

Search repository
Hiroshi Yoshitomi

× Hiroshi Yoshitomi

Hiroshi Yoshitomi

Search repository
Moritake Takashi

× Moritake Takashi

Moritake Takashi

Search repository
Kunugita Naoki

× Kunugita Naoki

Kunugita Naoki

Search repository
抄録
内容記述タイプ Abstract
内容記述 We demonstrate a practical calibration method and its applicability for a commercially available radiophotoluminescence dosemeter (RPLD), i.e., the GD-352M (AGC Techno Glass, Shizuoka, Japan) to eye lens dose monitoring, by performing the calibration according to the ISO recommendations. The calibration was then verified through a series of experiments. For verification of the derived calibration factor (1.21 ± 0.04, k = 1) of the RPLD, we performed standard irradiations in the ISO narrow series X-ray reference fields and the simulation measurements in the actual radiation fields in a hospital. The TLD-based commercially available dosemeters, DOSIRISTM was also put on the ISO cylinder phantom and the RANDO phantom together with the GD-352M in the verification experiments. The personal dose equivalents Hp(3) obtained from the GD-352M and those obtained from the DOSIRISTM were in good agreement with each other. Our results demonstrate the proper calibration of a commercially available RPLD that is applicable to the additional monitoring of the lens of the eyes for medical staff.
書誌情報 Radiation Protection Dosimetry

巻 198, 号 17, p. 1303-1312, 発行日 2022-07
出版者
出版者 Oxford University Press
ISSN
収録物識別子タイプ ISSN
収録物識別子 0144-8420
DOI
識別子タイプ DOI
関連識別子 10.1093/rpd/ncac167
戻る
0
views
See details
Views

Versions

Ver.1 2025-08-15 01:22:34.209261
Show All versions

Share

Share
tweet

Cite as

Other

print

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX
  • ZIP

コミュニティ

確認

確認

確認


Powered by WEKO3


Powered by WEKO3