{"_buckets": {"deposit": "b00a3aae-63f3-45ce-9d7b-d040e16bfe0f"}, "_deposit": {"created_by": 1, "id": "82908", "owners": [1], "pid": {"revision_id": 0, "type": "depid", "value": "82908"}, "status": "published"}, "_oai": {"id": "oai:repo.qst.go.jp:00082908", "sets": ["1"]}, "author_link": ["1015835", "1015833", "1015836", "1015826", "1015832", "1015834", "1015827", "1015825", "1015831", "1015830", "1015824", "1015828", "1015829"], "item_8_biblio_info_7": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "2021-05", "bibliographicIssueDateType": "Issued"}, "bibliographicIssueNumber": "5", "bibliographicPageStart": "e0250667", "bibliographicVolumeNumber": "16", "bibliographic_titles": [{"bibliographic_title": "PloS one"}]}]}, "item_8_description_5": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "We evaluated the long-term stability of a newly developed viscoelastic phantom made of polyacrylamide (PAAm) gel for magnetic resonance elastography (MRE) and ultrasound-based shear-wave elastography (US SWE). The stiffness of the cylindrical phantom was measured at 0, 13 and 18 months. Storage and loss moduli were measured with MRE, and shear-wave speed (SWS) was measured with US SWE. Long-term stability was evaluated in accordance with the Quantitative Imaging Biomarker Alliance (QIBA) profiles for each modality. The initial storage and loss moduli of the phantom were 5.01±0.22 and 1.11±0.15 respectively, and SWS was 2.57±0.04 m/s. The weight of the phantom decreased by 0.6% over the 18 months. When measured with MRE, the stiffness of the phantom decreased and changes to the storage and loss moduli were -3.0% and -4.6% between 0 and 13 months, and -4.3% and 0.0% between 0 and 18 months. The US measurements found that SWS decreased by 2.4% over the first 13 months and 3.6% at 18 months. These changes were smaller than the tolerances specified in the QIBA profiles, so the viscoelastic PAAm gel phantom fulfilled the condition for long-term stability. This new phantom has the potential to be used as a quality assurance and quality control phantom for MRE and US SWE.", "subitem_description_type": "Abstract"}]}, "item_8_relation_13": {"attribute_name": "PubMed番号", "attribute_value_mlt": [{"subitem_relation_type_id": {"subitem_relation_type_id_text": "34019551", "subitem_relation_type_select": "PMID"}}]}, "item_8_relation_14": {"attribute_name": "DOI", "attribute_value_mlt": [{"subitem_relation_type_id": {"subitem_relation_type_id_text": "10.1371/journal.pone.0250667", "subitem_relation_type_select": "DOI"}}]}, "item_8_relation_17": {"attribute_name": "関連サイト", "attribute_value_mlt": [{"subitem_relation_type_id": {"subitem_relation_type_id_text": "https://journals.plos.org/plosone/article?id=10.1371/journal.pone.0250667", "subitem_relation_type_select": "URI"}}]}, "item_8_source_id_9": {"attribute_name": "ISSN", "attribute_value_mlt": [{"subitem_source_identifier": "1932-6203", "subitem_source_identifier_type": "ISSN"}]}, "item_access_right": {"attribute_name": "アクセス権", "attribute_value_mlt": [{"subitem_access_right": "metadata only access", "subitem_access_right_uri": "http://purl.org/coar/access_right/c_14cb"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "Usumura, Masashi"}], "nameIdentifiers": [{"nameIdentifier": "1015824", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Kishimoto, Riwa"}], "nameIdentifiers": [{"nameIdentifier": "1015825", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Ishii, Koki"}], "nameIdentifiers": [{"nameIdentifier": "1015826", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Hotta, Eika"}], "nameIdentifiers": [{"nameIdentifier": "1015827", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Kershaw, Jeff"}], "nameIdentifiers": [{"nameIdentifier": "1015828", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Higashi, Tatsuya"}], "nameIdentifiers": [{"nameIdentifier": "1015829", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Obata, Takayuki"}], "nameIdentifiers": [{"nameIdentifier": "1015830", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Suga, Mikio"}], "nameIdentifiers": [{"nameIdentifier": "1015831", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Masashi, Usumura", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "1015832", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Riwa, Kishimoto", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "1015833", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Kershaw, Jeffrey", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "1015834", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Tatsuya, Higashi", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "1015835", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Takayuki, Obata", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "1015836", "nameIdentifierScheme": "WEKO"}]}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "eng"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "journal article", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "Longitudinal stability of a multimodal visco-elastic polyacrylamide gel phantom for magnetic resonance and ultrasound shear-wave elastography.", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "Longitudinal stability of a multimodal visco-elastic polyacrylamide gel phantom for magnetic resonance and ultrasound shear-wave elastography."}]}, "item_type_id": "8", "owner": "1", "path": ["1"], "permalink_uri": "https://repo.qst.go.jp/records/82908", "pubdate": {"attribute_name": "公開日", "attribute_value": "2021-06-02"}, "publish_date": "2021-06-02", "publish_status": "0", "recid": "82908", "relation": {}, "relation_version_is_last": true, "title": ["Longitudinal stability of a multimodal visco-elastic polyacrylamide gel phantom for magnetic resonance and ultrasound shear-wave elastography."], "weko_shared_id": -1}
Longitudinal stability of a multimodal visco-elastic polyacrylamide gel phantom for magnetic resonance and ultrasound shear-wave elastography.
https://repo.qst.go.jp/records/82908
https://repo.qst.go.jp/records/82908453717b3-1cbd-4c3d-a85a-106a7f7add07
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2021-06-02 | |||||
タイトル | ||||||
タイトル | Longitudinal stability of a multimodal visco-elastic polyacrylamide gel phantom for magnetic resonance and ultrasound shear-wave elastography. | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
アクセス権 | ||||||
アクセス権 | metadata only access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_14cb | |||||
著者 |
Usumura, Masashi
× Usumura, Masashi× Kishimoto, Riwa× Ishii, Koki× Hotta, Eika× Kershaw, Jeff× Higashi, Tatsuya× Obata, Takayuki× Suga, Mikio× Masashi, Usumura× Riwa, Kishimoto× Kershaw, Jeffrey× Tatsuya, Higashi× Takayuki, Obata |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | We evaluated the long-term stability of a newly developed viscoelastic phantom made of polyacrylamide (PAAm) gel for magnetic resonance elastography (MRE) and ultrasound-based shear-wave elastography (US SWE). The stiffness of the cylindrical phantom was measured at 0, 13 and 18 months. Storage and loss moduli were measured with MRE, and shear-wave speed (SWS) was measured with US SWE. Long-term stability was evaluated in accordance with the Quantitative Imaging Biomarker Alliance (QIBA) profiles for each modality. The initial storage and loss moduli of the phantom were 5.01±0.22 and 1.11±0.15 respectively, and SWS was 2.57±0.04 m/s. The weight of the phantom decreased by 0.6% over the 18 months. When measured with MRE, the stiffness of the phantom decreased and changes to the storage and loss moduli were -3.0% and -4.6% between 0 and 13 months, and -4.3% and 0.0% between 0 and 18 months. The US measurements found that SWS decreased by 2.4% over the first 13 months and 3.6% at 18 months. These changes were smaller than the tolerances specified in the QIBA profiles, so the viscoelastic PAAm gel phantom fulfilled the condition for long-term stability. This new phantom has the potential to be used as a quality assurance and quality control phantom for MRE and US SWE. | |||||
書誌情報 |
PloS one 巻 16, 号 5, p. e0250667, 発行日 2021-05 |
|||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 1932-6203 | |||||
PubMed番号 | ||||||
識別子タイプ | PMID | |||||
関連識別子 | 34019551 | |||||
DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1371/journal.pone.0250667 | |||||
関連サイト | ||||||
識別子タイプ | URI | |||||
関連識別子 | https://journals.plos.org/plosone/article?id=10.1371/journal.pone.0250667 |