量研学術機関リポジトリ「QST-Repository」は、国立研究開発法人 量子科学技術研究開発機構に所属する職員等が生み出した学術成果(学会誌発表論文、学会発表、研究開発報告書、特許等)を集積しインターネット上で広く公開するサービスです。 Welcome to QST-Repository where we accumulates and discloses the academic research results(Journal Publications, Conference presentation, Research and Development Report, Patent, etc.) of the members of National Institutes for Quantum Science and Technology.
Thank you very much for using our website. On the 11th of March 2019, this site was moved from our own network server to the JAIRO Cloud network server. If you previously bookmarked this site, that bookmark will no longer work. We would be grateful if you could bookmark the website again. Thank you very much for your understanding and cooperation.
The World Health Organisation (WHO) recommends that the concentration of radon in water should be no more than 100 kBq m−3 (100 BqL−1) and the Codex Alimentarius Commission states that the limit of quantification (LOQ) of a method should be no more than one-fifth of this value. In this study, a degassing method with an RAD7 device was used to measure radon concentrations in water, compared to a liquid scintillation counter (LSC) method used as the reference, to investigate whether the numerical value of the LOQ of this method was more than 1/5 (20 kBq m–3) of 100 kBq m–3. The degassing method with leak prevention was shown to reach a target value of 20 kBq m−3 or less under a relative humidity of 6% or lower in the chamber of the RAD7 device. Accordingly, the RAD7 degassing method with leak prevention can be used to accurately measure radon concentrations in water within the guidance level set out by the WHO.