@inproceedings{oai:repo.qst.go.jp:00075463, author = {YAMAGUCHI, Yuji and SANAMI, Toshiya and KOBA, Yusuke and UOZUMI, Yusuke and Koba, Yusuke}, book = {Proceedings of the Second International Symposium on Radiation Detectors and Their Uses (ISRD2018) JPS Conf. Proc.}, month = {Mar}, note = {We have developed a low threshold detector consisting of Bragg curve counter (BCC) and two built-in solid-state detectors (SSDs) to measure spectra of low energy charged particles emitted by nuclear reactions. Since the BCC, ionization chamber offers advantages of self particle identification capability and a few m-thick entrance window, the threshold energy less than 1 MeV/u is expected with particle identification. The detector is tested using 70 MeV protons for measurement of double-differential cross sections (DDXs) of charged particle production. Protons produced down to 1 MeV have been identified and resultant DXs for proton production have been obtained down to 1.5 MeV.}, pages = {011007-1--011007-6}, title = {Spectrum Measurement Down to 1 MeV/u Particles with Hydrogen-Identification Using Bragg Curve Counter}, volume = {24}, year = {2019} }