@article{oai:repo.qst.go.jp:00074759, author = {長谷, 純宏 and 佐藤, 勝也 and 千葉, 敦也 and 平野, 貴美 and Hase, Yoshihiro and Sato, Katsuya and Chiba, Atsuya and Hirano, Yoshimi}, journal = {QST Takasaki Annual Report 2017}, month = {Mar}, note = {The cluster ion-specific irradiation effects have been observed in the analysis of ion-material interaction due to their unique feature such as very high-density energy deposition. The cluster ion beams may also cause a unique irradiation effects in biological samples, however, hardly applied so far. This study is aimed to establish an irradiation method of cluster ions for bacteria and investigate the cluster ion-specific effects by comparing the irradiation effects of cluster and monomer ion beams. In order to evaluate the accuracy of the measurement of beam current using a Faraday cup, we measured the particle fluence using a solid-state track detector. The results suggested that the measurement of beam current with the Faraday cup equipped in the TC line is accurate with the possible errors of 3~4% in both cluster and monomer ion beams.}, title = {Evaluation of Particle Fluence of Cluster and Monomer Ion Beams Using a Solid-state Track Detector}, year = {2019} }