@misc{oai:repo.qst.go.jp:00071847, author = {Hara, Yousuke and Furukawa, Takuji and Mizushima, Kota and Saotome, Naoya and Saraya, Yuichi and Tansho, Ryohei and Takeshita, Eri and Shirai, Toshiyuki and Noda, Koji and 原 洋介 and 古川 卓司 and 水島 康太 and 早乙女 直也 and 皿谷 有一 and 丹正 亮平 and 竹下 英里 and 白井 敏之 and 野田 耕司}, month = {Nov}, note = {For quality assurance (QA) of therapeutic ion beams, QA tool having high spatial resolution and quick verification is required. The imaging system with a fluorescent screen is suitable for QA procedure. We developed a quick verification system (NQA-SCN) using a fluorescent screen with a charge-coupled device (CCD) camera for the sake of two dimensional dosimetry. The NQA-SCN is compact size to attach to irradiation port and water column. Several types of corrections were applied to the raw image obtained by the NQA-SCN. Our goal is to use the NQA-SCN for three-dimensional dose verification. In carbon-ion therapy, the fluorescent light is decreased by suffering from quenching effect due to the increased linear energy transfer (LET) in the Bragg peak. For the use of three-dimensional dose verification, as a first approach, we investigated the quenching effect of carbon-ion beam in water. Also, to evaluate the performance of NQA-SCN, we carried out experiments concerning QA procedures., 19th International conference on accelerators and beam utilizations (ICABU2015)}, title = {Development of Compact Dose Verification System Using a Fluorescent Screen for Carbon-ion Therapy}, year = {2015} }