{"created":"2023-05-15T14:51:56.392488+00:00","id":71001,"links":{},"metadata":{"_buckets":{"deposit":"03f6fc23-8f99-4cfd-a66b-f7b1788f2be5"},"_deposit":{"created_by":1,"id":"71001","owners":[1],"pid":{"revision_id":0,"type":"depid","value":"71001"},"status":"published"},"_oai":{"id":"oai:repo.qst.go.jp:00071001","sets":["10:28"]},"author_link":["697869","697870"],"item_10005_date_7":{"attribute_name":"発表年月日","attribute_value_mlt":[{"subitem_date_issued_datetime":"2012-12-07","subitem_date_issued_type":"Issued"}]},"item_10005_description_5":{"attribute_name":"抄録","attribute_value_mlt":[{"subitem_description":"An isocentric superconducting rotating-gantry for heavy-ion therapy is being designed. This rotating gantry can transport heavy ions having 430 MeV/u to an isocenter with irradiation angles over 0-360 degrees, and is further capable of performing the fast raster-scanning irradiation. For the magnets, combined-function superconducting-magnets were employed. The use of these superconducting magnets allowed us to design a compact gantry, while keeping a large scan size at the isocenter. The superconducting magnets were designed by using a 3D electromagnetic-field solver, the OPERA-3D code. Two superconducting magnets were manufactured, and magnetic field measurements are being performed. Results of the measurements were compared with those calculated with the 3D electromagnetic-field solver. We will present the design of the superconducting rotating-gantry and their combined-function superconducting magnets as well as results of the field measurements.","subitem_description_type":"Abstract"}]},"item_10005_description_6":{"attribute_name":"会議概要(会議名, 開催地, 会期, 主催者等)","attribute_value_mlt":[{"subitem_description":"XVI International Conference on Electromagnetic Isotope Separators and Techniques Related to their Applications (EMIS2012)","subitem_description_type":"Other"}]},"item_access_right":{"attribute_name":"アクセス権","attribute_value_mlt":[{"subitem_access_right":"metadata only access","subitem_access_right_uri":"http://purl.org/coar/access_right/c_14cb"}]},"item_creator":{"attribute_name":"著者","attribute_type":"creator","attribute_value_mlt":[{"creatorNames":[{"creatorName":"Iwata, Yoshiyuki"}],"nameIdentifiers":[{"nameIdentifier":"697869","nameIdentifierScheme":"WEKO"}]},{"creatorNames":[{"creatorName":"岩田 佳之","creatorNameLang":"en"}],"nameIdentifiers":[{"nameIdentifier":"697870","nameIdentifierScheme":"WEKO"}]}]},"item_language":{"attribute_name":"言語","attribute_value_mlt":[{"subitem_language":"eng"}]},"item_resource_type":{"attribute_name":"資源タイプ","attribute_value_mlt":[{"resourcetype":"conference object","resourceuri":"http://purl.org/coar/resource_type/c_c94f"}]},"item_title":"Development of a superconducting rotating-gantry and their combined-function superconducting magnets for heavy-ion therapy","item_titles":{"attribute_name":"タイトル","attribute_value_mlt":[{"subitem_title":"Development of a superconducting rotating-gantry and their combined-function superconducting magnets for heavy-ion therapy"}]},"item_type_id":"10005","owner":"1","path":["28"],"pubdate":{"attribute_name":"公開日","attribute_value":"2012-12-06"},"publish_date":"2012-12-06","publish_status":"0","recid":"71001","relation_version_is_last":true,"title":["Development of a superconducting rotating-gantry and their combined-function superconducting magnets for heavy-ion therapy"],"weko_creator_id":"1","weko_shared_id":-1},"updated":"2023-05-15T19:56:20.643208+00:00"}