@misc{oai:repo.qst.go.jp:00070959, author = {Tashima, Hideaki and Hirano, Yoshiyuki and Kinouchi, Shoko and Yoshida, Eiji and Ito, Hiroshi and Yamaya, Taiga and 田島 英朗 and 平野 祥之 and 木内 尚子 and 吉田 英治 and 伊藤 浩 and 山谷 泰賀}, month = {Nov}, note = {We are developing the OpenPET, an open type PET geometry having an accessible and observable open space between detector rings which allows easy access to a patient during PET measurements. Our first generation OpenPET which we called dual-ring OpenPET consisted of two detector rings having a gap between them and it could provide an extended axial field of view (FOV). For applications such as in-beam PET to monitor dose distribution in situ during particle therapy, however, higher sensitivity is required rather than a wide FOV. As the next generation OpenPET, we proposed the single-ring OpenPET which was a geometry having an accessible and observable open space with higher sensitivity and a reduced number of detectors than the earlier geometry. In this paper, we provide a theoretical analysis of the sensitivity for the single-ring OpenPET and we compare it with the dual-ring OpenPET and a geometry in which the conventional PET is positioned at a slant angle against the patient bed to form an accessible open space, which we called slant PET. The theoretical analysis shows that the single-ring OpenPET has a sensitivity 1.2 times higher than the dual-ring OpenPET and 1.3 times higher than the slant PET when designed for about 200 detector blocks, each with a front area of 2500 mm2. In addition, numerical simulation using the Geant4 Monte Carlo toolkit is done to show the three-dimensional imaging performance that the single-ring OpenPET realized with the ellipsoidal rings of block detectors. The numerical simulation demonstrates that the depth-of-interaction detector can provide uniform resolution even when the detectors are arranged in an ellipsoidal ring., 2012 Nuclear Science Symposium, Medical Imaging Conference & Workshop on Room-Temperature Semiconductor X-Ray and Gamma-Ray Detectors}, title = {Theoretical and Numerical Analysis of the Single-Ring OpenPET Geometry for In-beam PET}, year = {2012} }