{"created":"2023-05-15T14:47:17.335263+00:00","id":64847,"links":{},"metadata":{"_buckets":{"deposit":"fb930a03-d149-49b5-8b41-173af5379f82"},"_deposit":{"created_by":1,"id":"64847","owners":[1],"pid":{"revision_id":0,"type":"depid","value":"64847"},"status":"published"},"_oai":{"id":"oai:repo.qst.go.jp:00064847","sets":["10:29"]},"author_link":["639043","639048","639042","639047","639049","639041","639045","639044","639050","639046"],"item_10005_date_7":{"attribute_name":"発表年月日","attribute_value_mlt":[{"subitem_date_issued_datetime":"2012-11-25","subitem_date_issued_type":"Issued"}]},"item_10005_description_5":{"attribute_name":"抄録","attribute_value_mlt":[{"subitem_description":"目的 放射線医学総合研究所では、モニタユニット値を決定するために、水中を模擬した体系の拡大ブラックピーク中央深(以下、水SOBP中央深)線量を計算する方法が考えられた。今回、その計算方法に対し、照射野形状に依存したファントム内の散乱線が水SOBP中央深線量に与える影響を考慮し、精度向上を図った。\n方法 前立腺癌に主に使用する7種類の照射条件に対し、マルチリーフコリメータで形成した円形照射野サイズとレンジシフタ厚を変化させ、水SOBP中央深線量の変化をアイソセンター(以下、I.C)で測定した。それらの測定結果を多項式近似し、治療門に使用するレンジシフタ厚における照射野形状の水SOBP中央深線量を扇形積分により計算した。2011年4月5日〜2012年2月23日に治療室Cにおいて前述の照射条件で治療した前立腺患者の332門に対して、水SOBP中央深線量の計算値と実測値を比較した。\n結果 従来の方法による誤差(最大2.8%)が、線量評価点がI.Cの場合、最大0.6%、線量評価点がI.C以外の場合、最大-1.4%となった。\n結論 扇形積分を使った方法は線量評価点がI.Cの場合、臨床で許容できる精度を持つ。線量評価点がI.C以外の場合、軸外線量比等の補正による精度向上を今後検討する。","subitem_description_type":"Abstract"}]},"item_10005_description_6":{"attribute_name":"会議概要(会議名, 開催地, 会期, 主催者等)","attribute_value_mlt":[{"subitem_description":"日本放射線腫瘍学会第25回学術大会","subitem_description_type":"Other"}]},"item_access_right":{"attribute_name":"アクセス権","attribute_value_mlt":[{"subitem_access_right":"metadata only access","subitem_access_right_uri":"http://purl.org/coar/access_right/c_14cb"}]},"item_creator":{"attribute_name":"著者","attribute_type":"creator","attribute_value_mlt":[{"creatorNames":[{"creatorName":"田尻, 稔"}],"nameIdentifiers":[{"nameIdentifier":"639041","nameIdentifierScheme":"WEKO"}]},{"creatorNames":[{"creatorName":"前田, 貴雅"}],"nameIdentifiers":[{"nameIdentifier":"639042","nameIdentifierScheme":"WEKO"}]},{"creatorNames":[{"creatorName":"磯部, 喜治"}],"nameIdentifiers":[{"nameIdentifier":"639043","nameIdentifierScheme":"WEKO"}]},{"creatorNames":[{"creatorName":"黒岩, 俊隆"}],"nameIdentifiers":[{"nameIdentifier":"639044","nameIdentifierScheme":"WEKO"}]},{"creatorNames":[{"creatorName":"柴山, 晃一"}],"nameIdentifiers":[{"nameIdentifier":"639045","nameIdentifierScheme":"WEKO"}]},{"creatorNames":[{"creatorName":"田尻 稔","creatorNameLang":"en"}],"nameIdentifiers":[{"nameIdentifier":"639046","nameIdentifierScheme":"WEKO"}]},{"creatorNames":[{"creatorName":"前田 貴雅","creatorNameLang":"en"}],"nameIdentifiers":[{"nameIdentifier":"639047","nameIdentifierScheme":"WEKO"}]},{"creatorNames":[{"creatorName":"磯部 喜治","creatorNameLang":"en"}],"nameIdentifiers":[{"nameIdentifier":"639048","nameIdentifierScheme":"WEKO"}]},{"creatorNames":[{"creatorName":"黒岩 俊隆","creatorNameLang":"en"}],"nameIdentifiers":[{"nameIdentifier":"639049","nameIdentifierScheme":"WEKO"}]},{"creatorNames":[{"creatorName":"柴山 晃一","creatorNameLang":"en"}],"nameIdentifiers":[{"nameIdentifier":"639050","nameIdentifierScheme":"WEKO"}]}]},"item_language":{"attribute_name":"言語","attribute_value_mlt":[{"subitem_language":"jpn"}]},"item_resource_type":{"attribute_name":"資源タイプ","attribute_value_mlt":[{"resourcetype":"conference object","resourceuri":"http://purl.org/coar/resource_type/c_c94f"}]},"item_title":"ワブラー法を使った重粒子線治療におけるモニタユニット値の推定","item_titles":{"attribute_name":"タイトル","attribute_value_mlt":[{"subitem_title":"ワブラー法を使った重粒子線治療におけるモニタユニット値の推定"}]},"item_type_id":"10005","owner":"1","path":["29"],"pubdate":{"attribute_name":"公開日","attribute_value":"2012-11-25"},"publish_date":"2012-11-25","publish_status":"0","recid":"64847","relation_version_is_last":true,"title":["ワブラー法を使った重粒子線治療におけるモニタユニット値の推定"],"weko_creator_id":"1","weko_shared_id":-1},"updated":"2023-05-15T21:06:55.454524+00:00"}