@misc{oai:repo.qst.go.jp:00064290, author = {Ploc, Ondrej and Uchihori, Yukio and Kitamura, Hisashi and Kodaira, Satoshi and Yasuda, Nakahiro and Ambrozova, Iva and Mrazova, Zlata and Spurny, Frantisek and Ploc Ondrej and 内堀 幸夫 and 北村 尚 and 小平 聡 and 安田 仲宏 and Mrazova Zlata and Spurny Frantisek}, month = {Sep}, note = {Liulin and TEPCs are often used for measurements onboard spacecrafts. We did several meas-urements with TEPC Hawk and several Liulins in HIMAC in order to compare measured values, namely LET spectra, fluence, absorbed dose and dose equivalent. We observed a difference in distributions of events number in ADC channels measured with two identical Liulins irradiated by heavy ions under identical conditions. In order to measure dose quantities correctly, a recali-bration of Liulins was needed. We did it using PHITS. Further, we studied the effect of thick-nesses in front of active volume of different models of Liulin and Hawk. We calibrate Liulin to measure LET spectra in a broad parallel beam impinging detector under specified angle to get values comparable with Hawk's lineal energy spectra. Taken into account the different thick-nesses in front of active volumes, a good agreement between measurements with Liulin and Hawk in HIMAC can be stated., 15th WRMISS Workshop}, title = {Intercomparison measurements with Liulin and TEPC Hawk in HIMAC beams}, year = {2010} }