{"created":"2023-05-15T14:45:45.271536+00:00","id":62510,"links":{},"metadata":{"_buckets":{"deposit":"f55ba393-f3c0-4274-aefb-02fb762357c2"},"_deposit":{"created_by":1,"id":"62510","owners":[1],"pid":{"revision_id":0,"type":"depid","value":"62510"},"status":"published"},"_oai":{"id":"oai:repo.qst.go.jp:00062510","sets":["10:29"]},"author_link":["617557","617558","617560","617561","617559"],"item_10005_date_7":{"attribute_name":"発表年月日","attribute_value_mlt":[{"subitem_date_issued_datetime":"2008-04-06","subitem_date_issued_type":"Issued"}]},"item_10005_description_5":{"attribute_name":"抄録","attribute_value_mlt":[{"subitem_description":"X線CT装置の技術的発達による、診断能の向上、検査時間の短縮に伴い、成人だけでなく小児CT検査数も年々増加している。最近では、64列CT装置が臨床で使用され始めているが、最新型のX線CT装置を用いた成人および小児CT検査における国内の被ばく線量データは無いため、その実態解明が急務となっている。本研究では、64列CT装置を用いた成人および小児CT検査における被ばく線量評価を行った。\n 成人CT検査におけるスキャン範囲内の組織・臓器の臓器線量について、頭部CT検査では42- 55mGy、胸部CT検査では、8-35mGy、腹部−骨盤CT検査では12-33mGyであった。また、同様に小児CT検査について、頭部CT検査では、21-40mGy、胸部CT検査では、5-21mGy、腹部−骨盤CT検査では、4-14mGyで、各種CT検査における小児の被ばく線量は成人の20-50%程度に低減していた。本研究で評価された小児および成人CT検査における線量値は、CT検査における最適化の判断に有用と考えられる。","subitem_description_type":"Abstract"}]},"item_10005_description_6":{"attribute_name":"会議概要(会議名, 開催地, 会期, 主催者等)","attribute_value_mlt":[{"subitem_description":"第95回医学物理学会学術大会","subitem_description_type":"Other"}]},"item_access_right":{"attribute_name":"アクセス権","attribute_value_mlt":[{"subitem_access_right":"metadata only access","subitem_access_right_uri":"http://purl.org/coar/access_right/c_14cb"}]},"item_creator":{"attribute_name":"著者","attribute_type":"creator","attribute_value_mlt":[{"creatorNames":[{"creatorName":"藤井, 啓輔"}],"nameIdentifiers":[{"nameIdentifier":"617557","nameIdentifierScheme":"WEKO"}]},{"creatorNames":[{"creatorName":"青山, 隆彦"}],"nameIdentifiers":[{"nameIdentifier":"617558","nameIdentifierScheme":"WEKO"}]},{"creatorNames":[{"creatorName":"川浦, 稚代"}],"nameIdentifiers":[{"nameIdentifier":"617559","nameIdentifierScheme":"WEKO"}]},{"creatorNames":[{"creatorName":"小山, 修司"}],"nameIdentifiers":[{"nameIdentifier":"617560","nameIdentifierScheme":"WEKO"}]},{"creatorNames":[{"creatorName":"藤井 啓輔","creatorNameLang":"en"}],"nameIdentifiers":[{"nameIdentifier":"617561","nameIdentifierScheme":"WEKO"}]}]},"item_language":{"attribute_name":"言語","attribute_value_mlt":[{"subitem_language":"jpn"}]},"item_resource_type":{"attribute_name":"資源タイプ","attribute_value_mlt":[{"resourcetype":"conference object","resourceuri":"http://purl.org/coar/resource_type/c_c94f"}]},"item_title":"64列X線CT装置を用いた成人及び小児CT検査における被ばく線量評価","item_titles":{"attribute_name":"タイトル","attribute_value_mlt":[{"subitem_title":"64列X線CT装置を用いた成人及び小児CT検査における被ばく線量評価"}]},"item_type_id":"10005","owner":"1","path":["29"],"pubdate":{"attribute_name":"公開日","attribute_value":"2008-04-07"},"publish_date":"2008-04-07","publish_status":"0","recid":"62510","relation_version_is_last":true,"title":["64列X線CT装置を用いた成人及び小児CT検査における被ばく線量評価"],"weko_creator_id":"1","weko_shared_id":-1},"updated":"2023-05-15T21:29:58.300991+00:00"}