@article{oai:repo.qst.go.jp:00055963, author = {兼松, 伸幸 and 兼松 伸幸}, issue = {2}, journal = {医学物理}, month = {Sep}, note = {A guideline for quality assurance of CT systems for radiotherapy treatment planning is proposed. Quantitative interpretation of CT number is very important especially for range calculation in heavy charged particle radiotherapy, for which we adopted the polybinary calibration method to correct variations among CT systems and scanning conditions. Practical procedures for comissioning and constancy testing are documented along with the methodologies against various sources of uncertainty. We propose this guideline for quality assurance of heavy charged particle radiotherapy.}, pages = {140--146}, title = {多二元校正法による放射線治療計画用CT装置の品質管理}, volume = {23}, year = {2012} }