@article{oai:repo.qst.go.jp:00049014, author = {遠藤, 暁 and 佐藤, 斉 and takuro, shimazaki and kotani, kei and 須田, 充 and 濱野, 毅 and 梶本, 剛 and 田中, 憲一 and 星, 正治 and 遠藤 暁 and 佐藤 斉 and 須田 充 and 濱野 毅 and 梶本 剛 and 田中 憲一 and 星 正治}, issue = {3}, journal = {Radiat Environ Biophys}, month = {Jul}, note = {An LBO (Li2B4O7) walled ionization chamber was designed to monitor the epithermal neutron fluence in boron neutron capture therapy clinical irradiation. The thermal and epithermal neutron sensitivities of the device were evaluated using accelerator neutrons from the 9Be(d, n) reaction at a deuteron energy of 4 MeV (4 MeV d-Be neutrons). The response of the chamber in terms of the electric charge induced in the LBO chamber was compared with the thermal and epithermal neutron fluences measured using the gold-foil activation method. The thermal and epithermal neutron sensitivities obtained were expressed in units of pC cm2, i.e., from the chamber response divided by neutron fluence (cm-2). The measured LBO chamber sensitivities were 2.23 9 10-7 ± 0.34 9 10-7 (pC cm2) for thermal neutrons and 2.00 9 10-5 ± 0.12 9 10-5 (pC cm2) for epithermal neutrons. This shows that the LBO chamber is sufficiently sensitive to epithermal neutrons to be useful for epithermal neutron monitoring in BNCT irradiation.}, pages = {269--276}, title = {Determination of the thermal and epithermal neutron sensitivities of an LBO chamber}, volume = {56}, year = {2017} }