@article{oai:repo.qst.go.jp:00048564, author = {Koba, Yusuke and Shimomura, Risa and Chang, Weishan and Shinsho, Kiyomitsu and Yanagisawa, Shin and Wakabayashi, Genichro and Matsumoto, Kazuki and Ushiba, Hiroaki and Ando, Takayuki and Koba, Yusuke}, issue = {7}, journal = {Sensors and Materials}, month = {Oct}, note = {Cr doped alumina (Al2O3:Cr) ceramic TLD suitable to be a 2D imaging device in radiation therapy. In this work, we investigated the dose linearity in the dose level of radiotherapy and the LET dependence of Al2O3:Cr ceramic TLD using several HCP beams. Al2O3:Cr ceramic TLD had good linearity with the irradiated dose. But it had a large LET dependence and it was difficult to apply to dose measurement in HCP therapy directly. However, the combination of Al2O3:Cr ceramic TLD and Cr non-doped Al2O3 ceramic TLD may be expected to be applied to measurement of LET distribution using these different LET dependence.}, pages = {1599--1604}, title = {Dose Linearity and Linear Energy Transfer Dependence of Cr-doped Al2O3 Ceramic Thermoluminescence Detector}, volume = {30}, year = {2019} }