@article{oai:repo.qst.go.jp:00048351, author = {兼松, 伸幸 and 森, 慎一郎 and 稲庭, 拓 and 兼松 伸幸 and 森 慎一郎 and 稲庭 拓}, issue = {1}, journal = {医学物理}, month = {Sep}, note = {A calibration method for CT-number to stopping-power-ratio conversion was recently proposed as a revision of the Japanese de facto standard method that has been used at particle-therapy centers in Japan for over a decade. The revised method deals with 11 representative tissues with specific elemental composition and density, based on a latest compilation of standard tissue data. We report here how the revision was actually implemented into clinical practice. We applied the revised method to 7 CT-scanning conditions currently in use for treatment planning. For each condition, we derived CT numbers and stopping-power ratios of the representative tissues to constitute a polyline conversion function. We analyzed the target-depth change by the revision for 38 beams in treatment plans for 13 randomly sampled patients. The revision caused a mean change of +0.3 mm with a standard deviation of 0.4 mm. The maximum change was +1.2 mm or +0.5% of the depth, which may be clinically insignificant. The transition to the revised method was straightforward and would slightly improve the accuracy of beam range in particle therapy.}, pages = {34--37}, title = {粒子線治療計画のためのCT値阻止能比変換の校正方法の更新}, volume = {37}, year = {2017} }