@article{oai:repo.qst.go.jp:00047561, author = {Hirano, Yoshiyuki and Nishikido, Fumihiko and Kokuryo, Daisuke and Yamaya, Taiga and 平野 祥之 and 錦戸 文彦 and 國領 大介 and 山谷 泰賀}, issue = {2}, journal = {Radiological physics and technology}, month = {Jul}, note = {Multi-pixel photon counters (MPPCs) have been used instead of photomultiplier tubes for positron emission tomography combined with magnetic resonance (PET-MR). However, the effects of the magnetic field (MF) on the intrinsic properties-gain, cross-talk, after-pulsing, and dark-count-have not been sufficiently investigated. Therefore, we measured these properties for two types of MPPCs-S10931-50P and S12572-50P-in a static 7-T MF. These properties were measured with a pulse-shape analysis using pulse data acquired by a digital oscilloscope in the presence of the MF (w/MF) and the absence of the MF (w/o MF) by changing the supplied over-voltages (from 0.95 to 2.1 V for S10931 and from 2.1 to 3.3 V for S12572). No significant differences between the w/MF and w/o MF cases were observed for either MPPC, suggesting that the gain, cross-talk, after-pulsing, and dark-count are insensitive to a 7-T MF. The present work shows that constant MPPC performance is expected under a strong MF and demonstrates positive results for PET-MR.}, pages = {245--253}, title = {After-pulsing, cross-talk, dark-count, and gain of MPPC under 7-T static magnetic field.}, volume = {9}, year = {2016} }