@article{oai:repo.qst.go.jp:00045222, author = {Furukawa, Takuji and Saotome, Naoya and Inaniwa, Taku and Satou, Shinji and Noda, Kouji and Kanai, Tatsuaki and 古川 卓司 and 早乙女 直也 and 稲庭 拓 and 佐藤 眞二 and 野田 耕司 and 金井 達明}, issue = {6}, journal = {Medical Physics}, month = {May}, note = {The authors have developed a method to reconstruct the 3D dose distribution in a particle beam scanning irradiation system. In this scheme, 3D dose distribution is reconstructed by using the measured images of fluence distribution, which are taken for each iso-energy slice (i.e., the unit of the depth scanning). A fluorescent screen with a CCD camera is used to measure the fluence distribution. This system was installed at the HIMAC experimental port and tested by using carbon ion beams. Since a maximum difference between the reconstructed dose and the ionization chamber measurement was around 5% in the target volume, this system can be useful for quick verification of 3D dose distribution.}, pages = {2235--2242}, title = {Delivery verification using 3D dose reconstruction based on fluorescence measurement in a carbon beam scanning irradiation system}, volume = {35}, year = {2008} }