@article{oai:repo.qst.go.jp:00044748, author = {Honma, Toshihiro and Houjou, Satoru and Sakamoto, Yukio and Muramatsu, Masayuki and Miyahara, Nobuyuki and Yamada, Satoru and et.al and 本間 壽廣 and 北條 悟 and 坂本 幸雄 and 村松 正幸 and 宮原 信幸 and 山田 聰}, issue = {3}, journal = {Review of Scientific Instruments}, month = {Apr}, note = {A low-energy ion decelerator for the acceleration-deceleration system has been designed and installed in the beam-injection line of the NIRS-930 cyclotron in an attempt to increase the beam intensity from the cyclotron. With this acceleration-deceleration system, the beam intensity of C-12(4+) ions at the cyclotron exit has been increased by about five times. The system has an advantage in that it allows us to extract a high-current ion beam from an electron cyclotron resonance ion source because, independently of the injection-energy matching to the cyclotron, a large potential difference can be applied between the source and the extractor. The voltage applied to the extractor is about -12 kV, which is reduced to the ground potential by the decelerator. The electric field distributions in the deceleration system were carefully designed using a three-dimensional field simulator. Design studies of the decelerator as well as the result of a beam test are presented.}, pages = {03B909-1--03B909-3}, title = {Low-energy ion decelerator for an external injection line at the NIRS-930 cyclotron}, volume = {77}, year = {2006} }