@article{oai:repo.qst.go.jp:00044095, author = {松本, 政雄 and 岡本, 英明 and 黒田, 知純 and 鈴木, 隆一郎 and 松本, 徹 and その他 and 松本 政雄 and 松本 徹}, issue = {2}, journal = {医用画像情報学会雑誌}, month = {}, note = {We have attempted to calculate a primary X-ray spectrum of CT (Computed Tomography) system. Dose rates of primary X-ray spectrum are very high, so we reduced dose rate by using carbon scatterer and enabled to measure 90°-scattered spectrum with a CdZnTe detector. The primary X-ray spectrum can be calculated from measured data of 90°-scattered spectrum by counting backwards. In counting backwards, we used not only Klein-Nishina co-efficients but also response functions obtained by Monte Carlo methods, because Raylei scattering and multiple scattering maybe occur in carbon scatterer. In our results, the primary X-ray spectrum calculated from response functions obtained by Monte Carlo methods are roughly equivalent to that calculated from Klein-Nishina coefficients.}, pages = {88--96}, title = {CT装置の一次X線スペクトルの解析}, volume = {17}, year = {2000} }