@article{oai:repo.qst.go.jp:00044006, author = {Osuga, Toshiaki and Obata, Takayuki and Ikehira, Hiroo and Tanada, Shuji and Sasaki, Yasuhito and et.al and 大須賀 敏明 and 小畠 隆行 and 池平 博夫 and 棚田 修二 and 佐々木 康人}, issue = {10}, journal = {Artificial Organs}, month = {Oct}, note = {A procedure to determine dialysate pressure isobars by comparing the results of magnetic resonance imaging (MRI) and numerical simulation of the dialysate flow was established. The assumption used in the filtration analyses, that the dialysate pressure varies only in the central-axial direction, is supported by the finding that the determined isobars have no steep radial gradient. MRI of the contrast solution permeating the hollow fibers verifies the uniformity of the hollow fiber density in the dialyzer.}, pages = {907--909}, title = {Dialysate pressure isobars in a hollow-fiber dialyzer determined from magnetic resonance imaging and numerical simulation of dialysate flow.}, volume = {22}, year = {1998} }